17.Line-Related Sidebands Tests

Specification

Description

Equipment

Procedure

95 dB below the peak of a CW signal. Option $00: >75 dB below the peak of a CW signal.

The spectrally pure calibrator signal of the spectrum analyzer is applied to the analyzer input and the line related sidebands near the signal are measured.

None required

1.Press ~NSTR PRESET) on the analyzer. Connect CAL OUTPUT to

SIGNAL INPUT 2.

2.Press C-j 8 and adjust AMPTD CAL for a MARKER amplitude of -10.00 dBm f0.02 dB.

3.Press (~NSTR PRESET).

4.Key in the following analyzer settings:

 

 

 

 

 

 

.20

 

[CENTER FREQUENCY)

MHZ

 

 

 

 

 

 

 

-10 dBm

[

REFERENCE LEVEL] . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

 

 

] . . . .

. . . . . . . . . . .

. . . . . 600 Hz

CFREQUENCY S P A N

5.Wait for completion of sweep, then press MARKER

[ P E A K SEARCH),(jj).

6.Press (SHIFT) (-1 G, SWEEP @‘i7ZZ], 10 (j-/Iv to initiate video averaging of 10 sweeps. Wait for completion of sweeps.

7.Press MARKER (PEAK SEARCH],(~) and position movable marker at the peak of each line related sideband (120 Hz, 180 Hz, and 240 Hz for 60 Hz line frequency; 100 Hz, 150 Hz, and 200 Hz for 50 Hz line frequency, etc.). The MARKER A amplitude for each sideband should be c-85 dB (see Figure 2-31).

120 Hz (100

Hz)

 

dB

180 Hz (150

Hz)

 

dB

240 Hz (200

Hz)

 

 

dB

2-60 P e r f o r m a n c e T e s t s