25.Digital Storage Display Adjustments

Reference

Description

Equipment

Procedure

25. Digital Storage Display AdjustmentsA3A 1 TriggerA3A2 Intensity ControlA3A3 Line Generator

First, preliminary CRT graticule adjustments are performed to position the graticule on the CRT. These preliminary adjustments assume that repair has been performed on the associated circuitry. If no repair has been performed on the assemblies listed under REFERENCE, the preliminary adjustments are not necessary.

Next, the Sample and Hold Balance adjustments are performed. The horizontal and vertical Offset and Gain adjustments are performed, then the final CRT graticule adjustments are performed.

Last, the CRT annotation adjustments are performed to place the CRT annotation in proper location with respect to the CRT graticule.

DIGITAL VOLTMETER

Figure 3-89. Digital Storage Display Adjustments Setup

Digital Voltmeter (DVM) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP 3456A Digitizing Oscilloscope . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP 54501A

1.Place instrument upright as shown in Figure 3-89 with top and A3 Digital Storage cover removed.

2.Set LINE switch to ON and press [INSTR PRESET]
Preliminary Graticule Adjustments
3.Press TRACE A Cm].

4.Adjust A3A3R4 X GAIN and A3A3R5 Y GAIN to place graticule information completely on CRT. See Figure 3-90 for location of adjustment.

Adjustments 3-145