24.Track and Hold Adjustments

24.Adjust A3A9R52 GNEG for MARKER A level indication as indicated by CRT annotation of 100 ho.1 dB.

25.Repeat steps 4 through 24 until no further adjustments are required.

3-144 Adjustments