15. Spurious Responses Test

SPECTRUM ANALYZER

SYNTHESIZED SMEEPEA

10 DB

ATTENUATOR

300 MHZ LPF

J

Figure 2-25. Harmonic Distortion Test Setup

Note Equipment listed is for two test setups, Figure 2-25 and Figure 2-26.

Equipment

Synthesized

Sweeper

........... HP 8340A

 

Frequency

Synthesizer

H P 3335A

 

10 dB Attenuator (2 required)

: : ii< 84&A Opt 010

 

Lowpass Filter (300 MHz)

. . . . . . . HP 0955-0455

 

Adapter, Type N (m) to BNC (f) (2 required)

. . . . . . . . HP 1250-0780

 

Adapter, Type N (m) to SMA (f)

. . . . . . . HP 1250-1250

 

Adapter, Type N (f) to BNC (m)

. . . . . . . HP 1250-0077

 

Adapter, Type N (f) to BNC (f)

. . . . . . . HP 1250-1474

 

Directional

Bridge

. . . . . . . . . . . HP 8721A

 

Lowpass Filter (50 MHz) (2 required)

. . . . . . . HP 0955-0306

Procedure Harmonic Distortion

1.Connect equipment as shown in Figure 2-25.

2.On the spectrum analyzer, press [INSTR PRESET). Set the controls of the spectrum analyzer as follows:

CENTER FREQUENCY

.............. 280 MHz

FREQUENCY SPAN

. . . . . . . . . . . . . . . . 10 kHz

REFERENCE LEVEL

. . . . . . . . . . . . . - 20 dBm

3. On the synthesized sweeper, key in ~NSTR PRESET], Icw) 280 MHz,

( P O W E R L E V E L ) - 10 dBm.

4.On the spectrum analyzer, key in DISPLAY LINE CENTER] -90 dBm, MARKER [PEAK SEARCH] to position a marker on the peak of the displayed 280 MHz signal.

5.On the synthesized sweeper, press ~POWER LEVEL] and use the ENTRY knob to adjust the amplitude of the displayed 280 MHz

2-50 P e r f o r m a n c e T e s t s

Page 72
Image 72
HP 8568B manual Synthesized, Frequency, 3335A, Ii 84&A Opt, Lowpass Filter 50 MHz 2 required