4. Final Display Adjustments (SN 3004A and Above)

RI20R220R512R513 R319 R426 R437

Figure 3-32. Location of Final Display Adjustments on AlA

‘Ihble 3-6. Initial Adjustment Positions

AdjustmentPosition

AlA R120 Y GAIN

centered

AlA R127 Y POSN

centered

AlA R220 X GAIN

centered

AlA R227 X POSN

centered

AlA R319 INT GAIN

two-thirds clockwise

AlA R409 FOCUS COMP

centered

AlA R426 T/B FOC

centered

Al A2 R427 T/B CTR

centered

Al A2 R437 R/L FOC

centered

A lA2 R440 R/L CTR

centered

AlA R512 ORTHO

centered

AlA R513 3D

centered

AlA R516 INT LIM

fully counterclockwise

AlA R517 ASTIG

centered

Front-panel INTENSITY fully counterclockwise

Front-panel

FOCUS

centered

Front-panel

ALIGN

centered

4.Set the spectrum analyzer’s LINE switch to ON, and wait at least

5 minutes to allow the CRT and high-voltage circuits to warm up.

5.Set the front panel INTENSITY control fully counterclockwise and adjust AlA2R516 INT LIM until the display is just visable. See Figure 3-32.

6.Set the front-panel INTENSITY control fully clockwise.

7.Adjust AlA2R220 X GAIN, AlA2R227 X POSN, AlA2R120 Y GAIN, and AlA2R127 Y POSN for optimum centering of the display annotation and graticule pattern.

3 - 62 Adjustments