‘Ihble 3-2. Adjustable Components (continued)

Reference

Adjustment

Adjustment

Adjustment Function

Designator

Name

Number

 

AlA2R31

ORTHO

3

Sets orthogonality of CRT.

AlA2R32

PATTERN

3

Adjusts for optimum rectangular shape of CRT

 

 

 

display.

AlA2R35

INTENSITY

3

Sets adjustment range of front-panel INTENSITY

 

 

 

control.

 

LIMIT

 

 

AlA2R36

ASTIG

3

Adjusts astigmatism of CRT.

A lA2R30

FOCUS GAIN

4

Adjusts for optimum focus of CRT display.

AlA3R14

FOCUS LIMIT

3

Coarse adjusts CRT focus.

AlA4ClO

Cl0

3

Adjusts rise and fall times of X deflection amplifier

 

 

 

pulse.

AlA4Cll

Cl1

3

Adjusts rise and fall times of X deflection amplifier

 

 

 

pulse.

AlA4R7

X POSN

3

Adjusts horizontal position of trace.

AlA4R27

X GAIN

394

Adjusts horizontal gain of trace.

AlA4R28

HFGAIN

3

Adjusts rise and fall times or X deflection amplifier

 

 

 

pulse.

AlA5ClO

Cl0

3

Adjusts rise and fall times of Y deflection amplifier

 

 

 

pulse.

AlA5Cll

Cl1

3

Adjusts rise and fall times of Y deflection amplifier

 

 

 

pulse.

AlA5R7

Y POSN

394

Adjusts vertical position of trace.

AlA5R27

Y GAIN

394

Adjusts vertical gain of trace.

AlA5R28

HF GAIN

3,4

Adjusts rise and fall times of Y deflection amplifier

 

 

 

pulse.

AlA6R9

+ 15 SV ADJ

1

Adjusts + 15 V dc supply voltage.

AlA6R32

HV ADJUST

2

Adjusts CRT high voltage.

A3A8R9

FS

23

Adjusts high end of digitized sweep.

A3A8R14

ZERO

23

Adjusts low end of digitized sweep.

 

 

 

 

IF Serial Prefix 2637A and Below

 

 

 

 

 

A22

COARSE

12

Coarse-adjusts reference oscillator frequency.

A22

FINE

12

Fine-adjusts reference oscillator frequency.

3-12 Adjustments