12.Option 857

Amplitude Fidelity Test

Related Adjustment Log Amplifier Adjustments

Specification Log:

Incremental

fO.l dB/dB over 0 to 80 dB display

Cumulative

3 MHz to 30 Hz Resolution Bandwidth:

54~0.6 dB max over 0 to 70 dB display (20 - 30°C). sf1.5 dB max over 0 to 90 dB display

10 Hz Resolution Bandwidth:

sf0.8 dB max over 0 to 70 dB display (20 - 30°C). sk2.1 dB max over 0 to 90 dB display

Linear:

&3% of Reference Level for top 9-l/2 divisions of display

Description Amplitude fidelity in log and linear modes is tested by decreasing the signal level to the spectrum analyzer in 10 dB steps with a calibrated signal source and measuring the displayed amplitude change with the analyzer’s MARKER A function.

BNC TEE

Figure 5-1. Option 857 Amplitude Fidelity Test Setup

5-2 Option 857