17. Line-Related Sidebands TestsFigure 2-31. Line Related Sidebands MeasurementO p t i o n 4 0 0 1. Press ~NSTR PRESET]. Connect CAL OUTPUT to SIGNAL INPUT 2.

2.Press (ml 8 and adjust AMPTD CAL for a MARKER amplitude of -10. 00 dBm • tO.02 dB.

3.Press ONSTR PRESET).

4.Key in the following analyzer settings:

[

CENTER FREQUENCY

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .]

20 MHz

 

 

 

 

]

- 1 0 dBm

 

PREFERENCE L E V E L

 

 

 

 

 

3 kHz

(

FREQUENCY SPAN] . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

5. Wait for completion of the sweep, then press MARKER

 

 

 

 

 

[ P E A K SEARCH],(-).

 

6.Press (SHIFT) [VlDEo] G, SWEEP (-1, 10 cm/Iv to initiate video averaging of 10 sweeps. Wait for completion of sweeps.

7.Press MARKER SPEAK SEARCH],@ and position movable marker at the peak of each line related sideband (400 Hz, 800 Hz, and 1200 Hz). The MARKER A amplitude for each sideband should be c-75 dB.

400 Hz

 

 

dB

800 Hz

 

 

dB

1200 Hz

 

 

dB

Performance Tests 2 - 61