8. Frequency Response Test

42.Press DISPLAY LINE [ENTER] on the spectrum analyzer. Use the Display Line to measure the maximum and minimum points on the trace. (Disregard LO Feedthrough at 1 kHz.) Record the measurements below.

SIGNAL INPUT 1

(1 kHz to 100 kHz)

MaximumdBm

MinimumdBm

43.Set Display Line to peak of trace at 1 kHz.44.Key in the following spectrum analyzer settings:

TRACE A (CLEAR-WRITE)

. . . . . . . . . . . . . . . . . . . . .l kHz 1 kHz iO0 Hz

45. Set function generator controls as follows:

RANGE Hz . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100 (button) FREQUENCY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 MODULATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ..all out

46.Adjust function generator FREQUENCY as necessary to place signal near center graticule line and adjust AMPLITUDE VERNIER to place peak of signal at Display Line.

47.Key in the following spectrum analyzer settings:

FREQUENCY SPAN) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100 Hz &TT!, . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30 Hz

48.Set (CF STEP SIZE] to 100 Hz. Step spectrum analyzer

( C E N T E R FFIEQUENCY] from 1 kHz to 100 Hz with a, while setting function generator FREQUENCY to match spectrum analyzer center frequency at each step. Record level-at each setting.

SIGNAL INPUT 1

 

 

1000 Hz

 

 

dBm

900 Hz

dBm

800 Hz

 

dBm

700 Hz

 

dBm

600 Hz

 

dBm

500 Hz

 

dBm

400 Hz

 

dBm

300 Hz

 

dBm

200 Hz

 

dBm

100 Hz

 

dBm

P e r f o r m a n c e T e s t s 2 - 2 9