HP 8568B manual if Gain Uncertainty Measurement 2 dB

Models: 8568B

1 313
Download 313 pages 11.07 Kb
Page 58
Image 58

10. IF Gain Uncertainty Test

2 dB Gain Steps 13. Press QNSTR pREsETj,(jRECALL) 7.

14.Set [REFERENCE LEVEL] to -1.9 dBm.

15.Press MARKER (OFF). Set CVlDEo] to 100 Hz.

16.Set the frequency synthesizer for an output power level of -3.9 dBm. Set the amplitude increment for 2 dB steps.

17.Press MARKER [PEAK SEARCH),@

18.Set the analyzer CREFERENCE LEVEL) and the frequency synthesizer amplitude according to ‘Iable 2-13. At each setting, note the MKR A amplitude and record it in the table.

‘lhble 2-13. IF Gain Uncertainty, 2 dB Steps

 

Frequency

 

Deviation

LREFERENCE LEVEL)

 

CdBm)

 

Synthesizer

 

(MARKER A

 

 

 

Amplitude

 

‘Amplitude

 

 

 

Wm)

 

W-9

- 1 . 9

 

- 3 . 9

 

0 (ref)

- 3 . 9

 

- 5 . 9

 

 

- 5 . 9

 

- 7 . 9

 

 

- 7 . 9

 

- 9 . 9

 

 

- 9 . 9

 

- 11 . 9

 

 

 

 

 

 

 

 

YKRAE Hz

rp REF -1.0 dBm

AtfEN 10 dB

0.02 dB

1 m/

CENTER 22.001

03

Nbb

 

SPAM 2 . 0 0 kl4z

RES 81

1

hbix

VBU 100 Hz

SUP SW “.O

Figure 2-19. IF Gain Uncertainty Measurement (2 dB)

2 - 36 Performance Tests

Page 58
Image 58
HP 8568B manual if Gain Uncertainty Measurement 2 dB