7. 3 MHz Bandwidth Filter Adjustments

10 Hz Amplitude Adjustments

37.Connect CAL OUTPUT to RF INPUT. Key in [INSTR PRESET), [mj 9, (jEEki-- 10 Hz.

38.Adjust the instrument front panel FREQ ZERO control for maximum signal amplitude on the CRT display.

39. Key in CREs]U

s1 kHz andi DISPLAYn LINEg [ENTER)t . h

e

DATA knob, place the display line at the signal trace.

 

40.Key in [REs] 10 Hz.

41.Adjust the instrument front panel FREQ ZERO control for maximum signal amplitude on the CRT display.

42.Adjust A4A7R30 10 Hz AMPTD and A4A7R41 10 Hz AMPTD equal amounts to set the signal level the same as the reference level set in step 39. See Figure 3-38 for location of 10 Hz AMPTD adjusts.

43.Repeat steps 37 through 42 until no further adjustment is required.

3-76 Adjustments