‘Ihble 3-3. Factory-Selected Components (continued)

Reference Designator

A4A7R13

A4A7R23

A4A7R24

A4A7R34

A4A7R35

A4A7R45

A4A7R46

A4A7R56

A4A7R57

A4A7R60

A4A7R66

A4A7R68

A4A7R70

A4A7R72

A4A7R74

A4A7R76

A4A7R78

A4A7R80

A4A7R82

A4A7R84

A4A7R86

A4A7R88

A4A7R90

A4A7R92

A4A7R94

A4A7R96

A4A7R98

A4A7RlOO

A4A7R102

A4A7R104

Adjustment Range of Values

Procedure (0 or PF)

10.0 K to 17.8 K

10.0 K to 17.8 K

10.0 K to 17.8 K

10.0 K to 17.8 K

10.0 K to 17.8 K

10.0 K to 17.8 K

10.0 K to 17.8 K

7.50 K to 13.3 K

7.50 K to 13.3 K

1038.3 to 68.1

38.3to 68.1

100 to 178

383 to 681

1.47K to 2.61 K

38.3to 68.1

100 to 178

383 to 681

1.47K to 2.61 K

38.3to 68.1

100 to 178

383 to 681

1.47K to 2.61 K

3.83to 68.1

100 to 178

383 to 681

1.47 K to 2.61 K

3.83 to 68.1

100 to 178

383 to 681

1.47 K to 2.61 K

Function of Component

Adjusts crystal filter bandwidth. Adjusts crystal filter bandwidth. Adjusts crystal filter bandwidth. Adjusts crystal filter bandwidth. Adjusts crystal filter bandwidth. Adjusts crystal filter bandwidth. Adjusts crystal filter bandwidth. Adjusts crystal filter bandwidth. Adjusts crystal filter bandwidth.

Compensates for gain of A4A6Al. Adjusts crystal filter bandwidth. Adjusts crystal filter bandwidth. Adjusts crystal filter bandwidth. Adjusts crystal filter bandwidth. Adjusts crystal filter bandwidth. Adjusts crystal filter bandwidth. Adjusts crystal filter bandwidth. Adjusts crystal filter bandwidth. Adjusts crystal filter bandwidth. Adjusts crystal filter bandwidth. Adjusts crystal filter bandwidth. Adjusts crystal filter bandwidth. Adjusts crystal filter bandwidth. Adjusts crystal filter bandwidth. Adjusts crystal filter bandwidth. Adjusts crystal filter bandwidth. Adjusts crystal filter bandwidth. Adjusts crystal filter bandwidth. Adjusts crystal filter bandwidth. Adjusts crystal filter bandwidth.

A4A8C 14 A4A8C35 A4A8C43 A4A8C49 A4A8C78 A4A8C8 1 A4A8C82 A4A8C83

For Option 462, see back of thisit,able for exceptions to A4A7.

1.0to 8.2 Sets adjustment range of SYM.

180 to 270 Sets adjustment range of LC CTR.

1.0to 8.2 Sets adjustment range of SYM.

180 to 270 Sets adjustment range of LC CTR.

180 to 270 Sets adjustment range of LC CTR.

180 to 270 Sets adjustment range of LC CTR.

4 to 13 Sets adjustment range of center cap.

4 to 13 Sets adjustment range of center cap.

Adjustments 3 - 15

Page 129
Image 129
HP 8568B manual Adjustment Range of Values Procedure 0 or PF, To 17.8 K To 13.3 K 10 38.3 to 100 to 383 to To 2.61 K