21. Slope Compensation Adjustments

Note At this sweep time, some trace discontinuities are common.

14.Adjust A22R66 TILT for best flatness (clockwise rotation increases the power slope), and trigger two sweeps on the synthesized sweeper. See Figure 3-81 for the location of A22R66. Compare the resultant trace with the specification. Continue adjusting A22R66 until best flatness is achieved.

Note Best flatness is achieved when the maximum number of frequency points are on or near the -14 dBm reference.

A22

F R E O U E N C Y C O N T R O L

I/I IAII

R66 T I L T

Figure 3-81. Location of A22R66 TIIIT Adjustment

15.Press TRACE A, m, [PEAK SEARCH], and [MARKER DELTA). Using the data knob, place the marker on the lowest power peak. The marker’s absolute value should be less than 2 dB.

16.See Figure 3-82 for examples of typical displays of frequency response correctly and incorrectly adjusted.

FREQUENCY RE

L I M I T S

PROPERLY ADJi

MALAD.JUS TED

Figure 3-82. Slope Compensation Adjustment Waveforms

Adjustments 3-135