9.6 dB Resolution Bandwidth Adjustments

24.Press MARKER IOFF), then MARKER In].

25.Using the DATA knob, adjust the marker down one side of the displayed signal to the 6 dB point; CRT MKR annotation indicates

.500 x.

26.Adjust A4A9R65 10 kHz for MKR A indication of 5.00 kHz while maintaining the marker at .500 X using the DATA knob. Refer to Figure 4-7 for the adjustment location.

27.Press MARKER @. Adjust the marker to the 6 dB point on the opposite side of the signal (CRT MKR A annotation indicates 1.00 Xl.

28.The CRT MKR A annotation now indicates the 6 dB bandwidth of the 10 kHz bandwidth filter. The bandwidth should be 10.0 fl.O kHz

29.Key in (jj) 3 kHz, (FREQUENCY SPAN] 5 kHz, (PEAK SEARCH], and [MKRJ If necessary, readjust by pressing [REFERENCE LEVEL] and using the DATA knob to place the signal peak near the top of the graticule.

30.Press MARKER m and MARKER a.

31.Using the DATA knob, adjust the marker down one side of the displayed signal to the 6 dB point; CRT MKR A annotation indicates .500 X.

32.Adjust A4A9R66 3 kHz for MKR A indication of 1.5 kHz while maintaining the marker at .500 X using the DATA knob. Refer to Figure 4-7 for the adjustment location.

33.Press MARKER In]. Adjust the marker to the 6 dB point on the opposite side of the signal (CRT MKR A annotation indicates 1.00

Xl.

34.The CRT MKR (ZJ annotation now of the 3 kHz bandwidth filter. The f0.30 kHz

indicates the 6 dB bandwidth bandwidth should be 3.00