2. High-Voltage Adjustment (SN 3001A and Below)

DIGITIZING

OSCILLOSCOPE

HI-VOLTAGE

SIGNAL ANALYZER

Figure 3-5. High Voltage Adjustment Setup

Equipment Digital Voltmeter (DVM) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .HP 3456A DC High-Voltage Probe (1000: 1 divider) . . . . . . . . . . . . . . . . . . HP 34111A Display Adjustment PC Board (service accessory) . . . . . . .85662-60088 Digitizing Oscilloscope . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP 54501A 1O:l Divider Probe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP 10432A Function Generator (2 required) . . . . . . . . . . . . . . . . . . . . . . . . . . . HP 3312A

High-Voltage Adjustment Procedure

Warning

In the following procedure, it is necessary to probe voltages which, if contacted, could cause serious personal injury. Use a nonmetallic alignment tool when making adjustments. Be extremely careful.

Warning Do not attempt to measure the CRT filament voltage directly. The filament voltage is referenced to the high-voltage cathode and can only be measured safely with a special high-voltage true-rms voltmeter and probe.

1.Set the spectrum analyzer’s LINE switch to STANDBY.

2.Remove the top cover from the IF-Display Section, and connect the equipment as shown in Figure 3-5 and described in the following steps.

3.Set the DVM to the 100 V range, and connect the DVM to AlA7TP3 (+ 100 V). Do not use the high-voltage probe. See Figure 3-6 for the location of AlA7TP3.