9. Impulse Bandwidth Adjustments

[REFERENCE LEVEL] and using the DATA knob to place the signal peak near the top of the graticule.

24.Press MARKER IOFF), then MARKER (al.

25.Using the DATA knob, adjust the marker down one side of the displayed signal to the 7.3 dB point; CRT MKR annotation indicates 0.430 X.

26.Adjust A4A9R65 10 kHz for MKR A indication of 5.00 kHz while maintaining the marker at 0.430 X using the DATA knob. Refer to Figure 4-8 for the adjustment location.

27.Press MARKER Ia]. Adjust the marker to the 7.3 dB point on the opposite side of the signal (CRT MKR A annotation indicates 1.00 Xl*

28.The CRT MKR A annotation now indicates the impulse bandwidth of the 10 kHz bandwidth. The impulse bandwidth should be 10.0 fl . O kHz

29.Key in @E’GZT] 3 kHz, [FREQUENCY SPAN] 5 kHz, [PEAK SEARCH), and

@iiGGZQ. If necessary, readjust by pressing [REFERENCE LEVEL) and using the DATA knob to place the signal peak near the top of the graticule.

30.Press MARKER a and MARKER a.

31.Using the DATA knob, adjust the marker down one side of the displayed signal to the 7.3 dB point; CRT MKR A annotation indicates 0.430 X.

32.Adjust A4A9R66 3 kHz for MKR A indication of 1.5 kHz while maintaining the marker at 0.430 X using the DATA knob. Refer to Figure 4-8 for the adjustment location.

33.Press MARKER In]. Adjust the marker to the 7.3 dB point on the opposite side of the signal (CRT MKR A annotation indicates 1.00 Xl.

34.The CRT MKR @ annotation now indicates the impulse bandwidth of the 3 kHz bandwidth. The impulse bandwidth should be 3.00 ho.30 kHz

Option 462 4 - 29

Page 294
Image 294
HP 8568B manual Key in @E’GZT 3 kHz, Frequency Span 5 kHz, Peak SEARCH