2.Frequency Span Accuracy lkst

Related Adjustments Frequency Control Adjustments

50 MHz Voltage-Tuned Oscillator Adjustments

Specification

SpanUncertainty

>l MHz f(2% of the actual frequency separation between two points +0.5% of span setting)

51 MHz 445% of the actual frequency separation between two points +0.5% of span setting)

Description Frequency Span accuracy is determined by measuring a frequency at 5% of sweep and then at 95% of sweep. These frequencies correspond

to half a division from each edge of the CRT.

The spans chosen are based on the architecture of the HP 8568B RF hardware:

Span

Assembly Being Swept

2 0 0 H z

VT0

Oscillator

(low

divide)

100 kHz

VT0

Oscillator (low divide)

100.1 kHz

VT0

Oscillator

(high

divide)

1 MHz

VT0

Oscillator

(high

divide)

1.01 MHz

FM Coil of Yig Oscillator

20 MHz

FM Coil of Yig Oscillator

20.1 MHz

Main Coil of Yig Oscillator

1.5 GHz

Main Coil of Yig Oscillator

SYNTHESIZED WEEPER

Figure 2-3. Frequency Span Accuracy Test Setup

2-6 P e r f o r m a n c e T e s t s