4.Final Display Adjustments (SN 3004A and Above)

8.For an initial coarse focus, adjust the following potentiometers in the sequence listed:

AlA3R14 FOCUS LIMIT

AlA2R517 ASTIG

AlA2R513 3D

AlA2R409 FOCUS COMP

9.Press QNSTR PRESET), then adjust the reference level to bring the displayed noise to the top division of the graticule. Press CENTER dB/DIvj and key in 1 dB/DIV. The noise should now completely fill the CRT graticule pattern, illuminating a large rectangular area. If necessary, adjust the reference level until the graticule pattern is completely filled.

10.Press ISHIFT) (oFF)m and then C-1 loFFp to turn off the CRT annotation and graticule pattern.

Connect a 56503 photometer probe to the Tektronix J-16 digital photometer. Set the photometer to the Xl range.

11.Place the photometer light probe hood against the IF-Display Section glass RF1 filter, and adjust AlA2R319 INT GAIN for a photometer reading of 80 NITS (cd/m’).

This reading must be made with the glass RF1 filter in place in front of the CRT. It might be necessary to slightly trim the top and bottom of the photometer probe’s hood so that it will fit flush against the glass RF1 filter.

If a standard J-16 photometer is used (instead of metric option OZ), adjust AlA2R319 for a photometer reading of 23.5 fl (footlamberts).

12.Set the LINE switch to STANDBY and then back to ON. The spectrum analyzer power-up annotation should be visible on the CRT display. (This includes the firmware datecode.)

13.For the best focus near the center of the CRT display, adjust the following potentiometers in the sequence listed below. Repeat as needed to optimize center-screen focus.

AlA3R14 FOCUS LIMIT AlA2R517 ASTIG

AlA2R513 3D for best focus of annotation (short vectors) AlA2R409 FOCUS COMP for best focus of graticule lines (long vectors)

14.Adjust AlA2R426 T/B FOC for best focus at the top and bottom of the display.

15.Adjust AlA2R437 R/L FOC for best focus at the right and left sides of the display.

16.If the top and bottom (or right and left sides) of the display achieve best focus at different potentiometer settings, adjust AlA2R427 T/B CTR or AlA2R440 R/L CTR, and then readjust AlA2R426 T/B FOC or AlA2R437 R/L FOC to optimize overall focus.