Procedure

X and Y Deflection

Amplifier Pulse

Response Adjustments

Note

3.Preliminary Display Adjustments (SN 3004A and Above)

1.Connect a 1O:l (10 MQ) divider probe to the oscilloscope’s channel

1input and a 1O:l divider probe to the channel 4 input.

2.On the oscilloscope, press (RECALL) (CLEARI) to perform a soft reset.

3.On the oscilloscope, press (CHAN) more preset probe , select channel 1, and use the front-panel knob to select a 10: 1 probe.

4.Select channel 4, and use the front-panel knob to select a IO:1 probe.

5.Press ml.

6.Connect the channel 1 probe to the oscilloscope’s rear panel PROBE COMPENSATION AC CALIBRATOR OUTPUT connector. Press CAUTO- SCALE). Adjust the channel 1 probe for an optimum square wave display on the oscilloscope.

7.Connect the channel 4 probe to the oscilloscope’s rear panel PROBE COMPENSATION AC CALIBRATOR OUTPUT connector. Press [AUTO- SCALE]. Adjust the channel 4 probe for an optimum square wave display on the oscilloscope.

Each probe is now compensated for the oscilloscope input to which it is connected. Do not interchange probes without recompensating.

8.Connect the channel 1 IO:1 divider probe to AlA2TP204, and the channel 4 probe to AlA2TP205, as shown in Figure 3-23. Connect the probe ground leads to AlA2TP106. See Figure 3-24 and Figure 3-25 for the location of the assemblies and test points.

5On TERkllNATlON W

Figure 3-23. Preliminary Display Adjustments Setup

9.Remove the cover over A3 Digital Storage Section and remove A3A2 Intensity Control Assembly. Insert the Display Adjustment PC board (HP part number 85662-60088) into the A3A2 slot. See Figure 3-24 for the location of the A3A2 assembly.

Adjustments 3-53