13. 20 MHz Reference Adjustments

Press &K?@

Press CnTavj:

AVmarkers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ..on

Vmarker 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 800mv

Vmarker2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ..2.7V

start marker . . . . . . . . . . . . . . . . . place at 2.7V crossing

stop marker . . . . . . . . . . . . . place at next 2.7V crossing

2 1. Oscilloscope display should be a 10 MHz signal of TTL level; less than +0.8V to greater than +2.7V. See Figure 3-58 for a typical signal.

- 6 0 . 0 0 0 is

4 0 . 0 0 0 “S

1 4 0 . 0 0 0 ns

 

2 0 . 0 ns/dlv

 

1 f800.0mV

Figure 3-58. Typical Signal at A16TP3

22.Install Al6 20 MHz Reference without extenders and reconnect cable 7 (violet) to A16J3.

23.Connect power meter to front-panel CAL OUTPUT.

24.Adjust A26 CAL LEVEL A16R51 for power meter indication of -10.0 dBm f0.2 dB.

25.the A23A6 Comb Generator must be readjusted after adjusting the 20 MHz Reference. Refer to Adjustments 22, Comb Generator Adjustments, for adjustment procedure.

3-106 Adjustments