1. Center Frequency Readout Accuracy Test

 

 

 

 

 

 

 

 

 

 

 

Note

 

 

Spectrum analyzer

center frequency readout can fall outside

of

 

 

 

specified limits if 10 MHz frequency reference has not been calibrated

 

 

 

within the past year. To eliminate “frequency reference error x

 

 

 

tune frequency” term, substitute spectrum analyzer 10 MHz FREQ

 

 

 

REFERENCE rear panel output for frequency standard and repeat

 

 

 

test.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

‘Ihble 2-2. Center Frequency Readout Error Test Record

 

 

 

 

 

 

 

 

 

 

 

 

 

 

I

Spectrum Analyzer

 

 

 

 

 

 

 

 

 

 

 

Center Readout

 

 

 

[FREQUENCYSPAN)I(CENTER

FREQumcy~

 

 

 

 

 

 

 

 

(MHz)\I

 

 

 

 

(MHZ)\I

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Min

Measured

 

Max

 

 

100 MHz

100

 

98

 

 

102

 

 

100 MHz

500

 

498

 

 

502

 

 

100 MHz

1000

 

998

 

 

1002

 

 

 

 

 

 

 

 

 

 

 

10 MHz

100

 

99.8

 

 

100.2

 

 

10 MHz

500

 

499.8

 

 

500.2

 

 

10 MHz

1000

 

999.8

 

 

1000.2

 

 

10 MHz

1500

 

1499.8

 

 

1500.2

 

 

1 MHz

1000

 

999.98

 

 

1000.02

 

 

100 kHz

1000

 

999.998

 

 

1000.002

 

 

10 kHz

1000

 

999.9998

 

 

1000.0002

 

 

 

 

 

 

 

 

 

 

 

 

Performance Tests 2 - 5