11. Log Scale Switching Uncertainty Test

REP -9.8 dml

HKA

100.001

B MHZ

ATTEN 10 dedB

-a.a2

*em

b

2 dB/

/

I

\

CENTER 100.000

MHZ

VBW 1 0 0 kHz

SPAN

100 kHZ

RES BW

30 kHzkliz

SWP 20.0

nl*ec

Figure 2-20. Log Scale Switching Uncertainty Measurement

able 2-15. Log Scale Switching Uncertainty

SCALE (dB/DIV)

1

2

5

10

Allowable

Deviation

WV

0(ref)

f0.5

f0.5

f0.5