Model GFC7000TA Carbon Dioxide Analyzer

 

Table 3-1:

Display and Touchscreen Control Description

 

 

 

 

 

Description/Function

 

Field

 

 

 

 

 

 

 

 

 

 

 

Status

LEDs indicating the states of Sample, Calibration and Fault, as follows:

 

 

 

 

 

 

 

 

 

Name

Color

 

State

Definition

 

 

 

 

 

Off

Unit is not operating in sample mode, DAS is disabled.

 

 

 

 

 

On

Sample Mode active; Front Panel Display being updated; DAS data

 

 

SAMPLE

Green

 

 

being stored.

 

 

 

 

 

Blinking

Unit is operating in sample mode, front panel display being updated,

 

 

 

 

 

DAS hold-off mode is ON, DAS disabled

 

 

 

 

 

 

 

 

 

 

 

Off

Auto Cal disabled

 

 

CAL

Yellow

 

On

Auto Cal enabled

 

 

 

 

 

Blinking

Unit is in calibration mode

 

 

 

 

 

 

 

 

 

FAULT

Red

 

Off

No warnings exist

 

 

 

Blinking

Warnings exist

 

 

 

 

 

 

 

 

 

 

 

 

Conc

Displays the actual concentration of the sample gas currently being measured by the analyzer in the

 

currently selected units of measure

 

 

 

Mode

Displays the name of the analyzer’s current operating mode

 

Param

Displays a variety of informational messages such as warning messages, operational data, test function

 

values and response messages during interactive tasks.

 

 

 

 

 

 

Control Buttons

Displays dynamic, context sensitive labels on each button, which is blank when inactive until applicable.

 

Figure 3-3 shows how the front panel display is mapped to the menu charts illustrated in this manual. The Mode, Param (parameters), and Conc (gas concentration) fields in the display screen are represented across the top row of each menu chart. The eight touchscreen control buttons along the bottom of the display screen are represented in the bottom row of each menu chart.