Teledyne gfc 7000ta operation manual Glossary, Term Description/Definition

Models: gfc 7000ta

1 280
Download 280 pages 4.46 Kb
Page 64
Image 64

FAQ

Model GFC7000TA Carbon Dioxide Analyzer

Q:Where is the sintered filter/sample flow control orifice?

A: These components are located inside the flow control assembly that is attached to the inlet side of the sample pump, see Figure 3-3. See Section 11.6.1 for instructions on disassembly and replacement.

Q:How do I set up a SEQUENCE to run a nightly calibration check? A: The setup of this option is located in Section 7.6.

Q:How do I set the analog output signal range and offset?

A: Instructions for this can be found in Section 6.13.4 which describes analog I/O configuration.

Q:What is the averaging time for an GFC 7000TA?

A: The default averaging time, optimized for ambient pollution monitoring, is 150 seconds for stable concentrations and 10 seconds for rapidly changing concentrations; See Section10.5.1 for more information. However, it is adjustable over a range of 0.5 second to 200 seconds (please contact customer service for more information).

4.2. Glossary

Note: Some terms in this glossary may not occur elsewhere in this manual.

Term

Description/Definition

10BaseT

an Ethernet standard that uses twisted (“T”) pairs of copper wires to transmit at

 

10 megabits per second (Mbps)

100BaseT

same as 10BaseT except ten times faster (100 Mbps)

 

 

APICOM

name of a remote control program offered by Teledyne to its customers

 

 

ASSY

Assembly

 

 

CAS

Code-Activated Switch

 

 

CD

Corona Discharge, a frequently luminous discharge, at the surface of a

 

conductor or between two conductors of the same transmission line,

 

accompanied by ionization of the surrounding atmosphere and often by a power

 

loss

CE

Converter Efficiency, the percentage of light energy that is actually converted

 

into electricity

CEM

Continuous Emission Monitoring

 

 

Chemical formulas that may be included in this document:

CO2

carbon dioxide

C3H8

propane

CH4

methane

H2O

water vapor

HC

general abbreviation for hydrocarbon

Teledyne Analytical Instruments

46

Page 64
Image 64
Teledyne gfc 7000ta operation manual Glossary, Term Description/Definition