Troubleshooting

Model GFC7000TA Carbon Dioxide Analyzer

Table 10-2: Test Functions - Indicated Failures

TEST

FUNCTIONS

TIME

RANGE

STABIL

CO2 MEAS

&

CO2 REF

MR RATIO

PRES

SAMPLE FL

SAMPLE

TEMP

BENCH TEMP

WHEEL TEMP

BOX TEMP

PHT DRIVE

INDICATED FAILURE(S)

Time of day clock is too fast or slow: To adjust See Section 6.6. Battery in clock chip on CPU board may be dead.

Incorrectly configured measurement range(s) could cause response problems with a Data logger or chart recorder attached to one of the analog output.

If the Range selected is too small, the recording device will over range.

If the Range is too big, the device will show minimal or no apparent change in readings.

Indicates noise level of instrument or CO2 concentration of sample gas (See Section 10.4.2 for causes).

If the value displayed is too high the IR Source has become brighter. Adjust the variable gain potentiometer on the sync/demod board

If the value displayed is too low or constantly changing and the CO2 REF is OK:

oFailed multiplexer on the mother board o Failed sync/demod board

o Loose connector or wiring on sync/demod board

Flow of purge gas to the GFC wheel housing may have stopped

If the value displayed is too low or constantly changing and the CO2 REF is BAD:

oGFC wheel stopped or rotation is too slow o Failed sync/demod board IR source

o Failed IR source o Failed relay board o Failed I2C bus

o Failed IR photo-detector

When the analyzer is sampling zero air and the ratio is too low:

oThe reference cell of the GFC wheel is contaminated or leaking.

oThe alignment between the GFC wheel and the segment sensor, the M/R sensor or both is

incorrect.

oFailed sync/demod board

oFlow of purge gas to the GFC wheel housing may have stopped When the analyzer is sampling zero air and the ratio is too high:

o Zero air is contaminated o Failed IR photo-detector

See Table 11-1 for SAMPLE PRES WARN

Check for gas flow problems. see Section 11.1.6

SAMPLE TEMP should be close to BENCH TEMP. Temperatures outside of the specified range or oscillating temperatures are cause for concern

Bench temp control improves instrument noise, stability and drift. Temperatures outside of the specified range or oscillating temperatures are cause for concern. See Table 11-1 for BENCH TEMP WARNING Wheel temp control improves instrument noise, stability and drift. Outside of set point or oscillating temperatures are cause for concern. See Table 11-1 for WHEEL TEMP WARNING

If the box temperature is out of range, check fan in the power supply module. Areas to the side and rear of instrument should allow adequate ventilation. See Table 11-1 for BOX TEMP WARNING.

If this drive voltage is out of range it may indicate one of several problems:

-A poor mechanical connection between the various components in inside the detector housing

-An electronic failure of the IR Photo-Detector’s built-in cooling circuitry, or;

-A temperature problem inside the analyzer chassis. In this case other temperature warnings would also be active such as BENCH TEMP WARNING or BOX TEMP WARNING.

Teledyne Analytical Instruments

222

Page 240
Image 240
Teledyne gfc 7000ta operation manual Test Functions Indicated Failures

gfc 7000ta specifications

The Teledyne GFC 7000TA is a cutting-edge, high-performance gas flow calibrator designed for both laboratory and field applications. This instrument is renowned for its accuracy, versatility, and reliability, making it an ideal choice for various industries, including aerospace, environmental monitoring, and manufacturing.

One of the standout features of the GFC 7000TA is its advanced measurement technology. It utilizes state-of-the-art thermal mass flow meters, which allow for precise fluid measurement across a wide range of flow rates. This technology ensures highly accurate readings, which are essential for applications requiring stringent compliance with regulatory standards.

The GFC 7000TA is capable of accommodating various types of gases, making it particularly versatile. Users can calibrate flow for gases such as air, nitrogen, or carbon dioxide, providing flexibility in a multitude of applications. The device also supports dual-channel capabilities, allowing simultaneous measurement and calibration of two different gas types, enhancing productivity and efficiency in processes.

Another notable characteristic of the GFC 7000TA is its user-friendly interface. Equipped with a high-resolution touchscreen display, the device offers intuitive navigation and data visualization. Users can easily set parameters, view real-time measurements, and generate printed reports. This interface reduces the learning curve for new operators, allowing organizations to implement the GFC 7000TA with minimal training.

In addition to its advanced measuring capabilities, the GFC 7000TA features a robust design for durability in harsh environments. The instrument is compliant with various international standards, ensuring its reliability and performance in various conditions. Its portable form factor makes it convenient for fieldwork, while the rugged construction provides peace of mind for users working in demanding settings.

The GFC 7000TA also incorporates modern connectivity options. With built-in USB and Ethernet ports, users can easily integrate the device into their existing systems for data logging and remote monitoring. This connectivity feature enables real-time data access and enhances the potential for analysis and optimization of gas flow systems.

In summary, the Teledyne GFC 7000TA is a high-performance gas flow calibrator characterized by its advanced thermal mass flow measurement technology, versatility in accommodating various gases, user-friendly interface, and robust design. With its modern connectivity features, this instrument stands out as a vital tool for professionals seeking precision and reliability in gas flow applications.