Operating Instructions

Model GFC7000TA Carbon Dioxide Analyzer

6.2.1. Test Functions

Test functions are shown on the front panel display while the analyzer is in SAMPLE mode. They provide information about the present operating status of the instrument and can be used troubleshooting and can be recorded in one of the DAS channels (Section 6.7) for data analysis. To view, press the <TST TST> buttons.

Table 6-2: Test Functions Defined

PARAMETER

DISPLAY

UNITS

MEANING

Range

RANGE

PPB, PPM,

The full scale limit at which the reporting range of the analyzer is

 

RANGE11

currently set. THIS IS NOT the Physical Range of the instrument.

 

UGM, MGM

 

RANGE21

See Section 6.8.1 for more information.

 

 

O2 Range

O2 RANGE

%

The range setting for the O2 Sensor

Stability

STABIL

PPB, PPM

Standard deviation of CO2 concentration readings. Data points are

 

UGM, MGM

recorded every ten seconds using the last 25 data points.

 

 

CO2 Measure

MEAS

MV

The demodulated, peak IR detector output during the measure

 

portion of the GFC Wheel cycle.

 

 

 

CO2 Reference

REF

MV

The demodulated, peak IR detector output during the reference

 

portion of the GFC wheel cycle.

 

 

 

Measurement /

MR Ratio

 

The result of CO2 MEAS divided by CO2 REF. This ratio is the

Reference Ratio

-

primary value used to compute CO2 concentration. The value

 

 

 

displayed is not linearized.

Sample Pressure

PRES

In-Hg-A

The absolute pressure of the Sample gas as measured by a

 

pressure sensor located inside the sample chamber.

 

 

 

Sample Flow

SAMPLE FL

 

Sample mass flow rate. This is computed from the differential

 

cc/min

between the pressures measured up-stream and down-stream of the

 

 

 

sample critical flow orifice pressures.

Sample

SAMP TEMP

°C

The temperature of the gas inside the sample chamber.

Temperature

 

 

 

Bench

BENCH TEMP

°C

Optical bench temperature.

Temperature

 

 

 

Wheel

WHEEL TEMP

°C

Filter wheel temperature.

Temperature

 

 

 

Box Temperature

BOX TEMP

°C

The temperature inside the analyzer chassis.

O2 Cell

O2 CELL

°C

O2 sensor cell temperature.

Temperature2

TEMP2

Photo-detector

PHT DRIVE

 

The drive voltage being supplied to the thermoelectric coolers of the

Temp. Control

mV

IR photo-detector by the sync/demod Board.

Voltage

 

 

 

 

 

Slope

SLOPE

 

The sensitivity of the instrument as calculated during the last

 

-

calibration activity. The SLOPE parameter is used to set the span

 

 

 

calibration point of the analyzer.

Offset

 

 

The overall offset of the instrument as calculated during the last

 

OFFSET

-

calibration activity. The OFFSET parameter is used to set the zero

 

 

 

point of the analyzer response.

O2 Sensor Slope 2

O2 SLOPE

-

O2 slope, computed during zero/span calibration.

O2 Sensor Offset 2

O2 OFFSET

-

O2 offset, computed during zero/span calibration.

Test channel

TEST

mV, mA

Displays the signal level of the TEST analog output channel. Only

output signal

appears when the TEST channel has been activated.

 

 

Current Time

TIME

-

The current time. This is used to create a time stamp on DAS

 

readings, and by the AUTOCAL feature to trigger calibration events.

 

 

 

1Only appears when the instrument’s reporting range mode is set for DUAL or AUTO

2Only appears when the optional O2 sensor is installed.

Teledyne Analytical Instruments

65

Page 83
Image 83