Troubleshooting

Model GFC7000TA Carbon Dioxide Analyzer

 

 

 

 

JP4 Connector to Opto-Pickup

D1 – M/R Sensor Status

 

 

Board

 

 

 

 

D2 – Segment Sensor Status

 

 

 

 

 

Figure 10-4: Sync/Demod Board Status LED Locations

10.1.4.3. Relay Board Status LEDs

There are eight LEDs located on the Relay Board. The most important of which is D1 is the most important LED, which indicates the health of the I2C bus. If D1 is blinking the other faults following LEDs can be used in conjunction with DIAG menu signal I/O to identify hardware failures of the relays and switches on the relay (See Section 6.9.2 and Appendix D).

Table 10-4: I2C Status LED Failure Indications

LED

FUNCTION

FAULT STATUS

INDICATED FAILURE(S)

D1

I2C bus Health

Continuously ON

Failed/Halted CPU

(Watchdog

or

Faulty Motherboard, or Relay Board

(Red)

Faulty Connectors/Wiring to/from Relay Board

Circuit)

Continuously OFF

 

Failed/Faulty +5 VDC Power Supply (PS1)

 

 

 

STATUS LED’s

DC VOLTAGE TEST

POINTS

 

RELAY PCA

Figure 10-5: Relay Board Status LEDs