Teledyne gfc 7000ta Analog Output Step Test, To enter the signal I/O test mode, press

Models: gfc 7000ta

1 280
Download 280 pages 4.46 Kb
Page 140
Image 140

Operating Instructions

Model GFC7000TA Carbon Dioxide Analyzer

To enter the signal I/O test mode, press:

SAMPLE

RANGE = 500.000 PPM

CO2 =X.XXX

< TST TST > CAL

SETUP

 

 

 

DIAG

SIGNAL I / O

 

Use NEXT & PREV to

 

 

 

move between signal

PREV NEXT JUMP

 

ENTR EXIT

types.

 

 

 

 

 

 

EXIT returns to the main SAMPLE display

SETUP X.X

PRIMARY SETUP MENU

 

CFG DAS RNGE PASS CLK MORE

EXIT

 

 

 

 

 

 

SETUP X.X

SECONDARY SETUP MENU

 

DIAG I / O Test Signals Displayed Here

PREV NEXT JUMP

PRNT EXIT

 

 

EXAMPLE

 

 

 

DIAG I / O

JUMP TO: 12

Use JUMP to go directly to a specific signal location

See Appendix A-4 for a complete list of available SIGNALS

EXAMPLE:

COMM VARS DIAG ALRM

EXIT

1 2

ENTR EXIT

Enter 12 to Jump to

12) ST_CONC_VALID

SAMPLE

 

ENTER SETUP PASS: 818

 

8

1

8

ENTR EXIT

 

 

 

 

DIAG I / O

ST_CONC_VALID = ON

 

Exit to return

 

 

 

 

to the

PREV NEXT JUMP

ON

PRNT EXIT

DIAG menu

 

 

 

 

 

 

Pressing PRNT will send a formatted printout to the serial port and can be captured

with a computer or other output device.

6.13.3. Analog Output Step Test

This test can be used to check the accuracy and proper operation of the analog outputs. The test forces all four analog output channels to produce signals ranging from 0% to 100% of the full scale range in 20% increments. This test is useful to verify the operation of the data logging/recording devices attached to the analyzer.