Image 143

Operating Instructions

Model GFC7000TA Carbon Dioxide Analyzer

The following DC current output limits apply to the current loop modules:

Table 6-18: Analog Output Current Loop Range

RANGE

MINIMUM OUTPUT

MAXIMUM OUTPUT

0-20 mA

0 mA

20 mA

These are the physical limits of the current loop modules, typical applications use 2-20 or 4-20 mA for the lower and upper limits. Please specify desired range when ordering this option.

The default offset for all ranges is 0 mA.

Pin assignments for the output connector at the rear panel of the instrument are shown in Table 6-19.

 

 

ANALOG OUT

 

 

 

A1

A2

A3

A4

+ -

+ - +

-

+ -

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Table 6-19: Analog Output Pin Assignments

PIN

ANALOG

VOLTAGE

CURRENT

 

OUTPUT

SIGNAL

SIGNAL

1

A1

V Out

I Out +

2

Ground

I Out -

 

 

 

 

 

3

A2

V Out

I Out +

4

Ground

I Out -

 

5

A31

V Out

I Out +

6

Ground

I Out -

 

7

A4

V Out

Not used

8

Ground

Not used

 

1Output A3 is only used when the O2 sensor option is installed

See Figure 3-2 for the location of the analog output connector on the instrument’s rear panel.