Theory of Operation

Model GFC7000TA Carbon Dioxide Analyzer

The four sample and hold circuits are:

DESIGNATION

ACTIVE WHEN:

 

IR BEAM PASSING THROUGH

SEGMENT SENSOR PULSE IS:

Measure Gate

MEASUREMENT cell of GFC Wheel

HIGH

Measure Dark Gate

MEASUREMENT cell of GFC Wheel

LOW

 

 

 

Reference Gate

REFERENCE cell of GFC Wheel

HIGH

 

 

 

Reference Dark Gate

REFERENCE cell of GFC Wheel

LOW

 

 

 

Timing for activating the Sample and Hold circuits is provided by a phase lock loop circuit (PLL). Using the segment sensor output as a reference signal the PLL generates clock signal at ten times that frequency. This faster clock signal is used by the PLD to make the sample and hold circuits capture the signal during the center portions of the detected waveform, ignore the rising and falling edges of the detector signal.

Sample & Hold

Active

Detector

Output

Sample & Hold

Inactive

Figure 9-14: Sample & Hold Timing

9.3.4.3. Sync/Demod Status LED’s

The following two status LED’s located on the synch/demod board provide additional diagnostic tools for checking the GFC wheel rotation.

Table 9-1: Sync/Demod Status LED Activity

LED

FUNCTION

STATUS OK

FAULT STATUS

D1

M/R Sensor Status

LED flashes approximately

LED is stuck

 

 

2/second

ON or OFF

D2

Segment Sensor

LED flashes approximately

LED is stuck

 

Status

6/second

ON or OFF

See Section 11.1.4 for more information.