Operating Instructions

Model GFC7000TA Carbon Dioxide Analyzer

6.13.5. Electric Test

The electric test function substitutes simulated signals for CO2 MEAS and CO2 REF, generated by circuitry on the sync/demod board, for the output of the IR photo-detector. While in this mode the user can also view the same test functions viewable from the main SAMPLE display. When the test is running, the concentration reported on the front panel display should be 40.0 ppm.

SAMPLE

RANGE = 500.000 PPM

CO2 =X.XXX

< TST TST > CAL

SETUP

 

 

 

 

 

 

SETUP X.X

PRIMARY SETUP MENU

 

CFG DAS RNGE PASS CLK MORE

EXIT

 

 

 

 

SETUP X.X SECONDARY SETUP MENU

COMM VARS DIAG ALRM

EXIT

 

 

 

SETUP X.X

ENTER DIAG PASS: 818

8 1 8

ENTR EXIT

 

 

DIAG

SIGNAL I / O

 

 

NEXT

ENTR

EXIT

 

 

 

Repeat Pressing NEXT unti . . .

 

 

 

 

DIAG

ELECTRIC TEST

 

 

PREV NEXT

ENTR

EXIT

 

 

 

 

 

 

 

 

DIAG ELEC

RANGE=50.000 PPM

CO2= 40.0

 

<TST TST>

 

EXIT

 

 

 

 

 

 

 

 

 

Press <TST TST> to view Test Functions

NOTE: CO2 MEAS and CO2 REF will be artificially altered

to enforce a CO2 reading of 40.0 ppm.

All other Test Functions will report the correct operational

value

Exit returns to the

DIAG Menu