Operating Instructions

Model GFC7000TA Carbon Dioxide Analyzer

6.8.7. Dilution Ratio

The dilution ratio is a software option that allows the user to compensate for any dilution of the sample gas before it enters the sample inlet. Using the dilution ratio option is a 4-step process:

1.Select reporting range units: Follow the procedure in Section 6.8.6.

2.Select the range: Use the procedures in Sections 6.8.2 – 6.8.5. Make sure that the SPAN value entered is the maximum expected concentration of the undiluted calibration gas and that the span gas is either supplied through the same dilution inlet system as the sample gas or has an appropriately lower actual concentration. For example, with a dilution set to 100, a 10 ppm gas can be used to calibrate a 1000 ppm sample gas if the span gas is not routed through the dilution system. On the other hand, if a 1000 ppm span gas is used, it needs to pass through the same dilution steps as the sample gas.

3.Set the dilution factor as a gain (e.g., a value of 20 means 20 parts diluting gas and 1 part of sample gas):

SAMPLE

RANGE = 500.000 PPM

CO2 =X.XXX

< TST TST > CAL

SETUP

 

 

 

 

SETUP C.3

 

PRIMARY SETUP MENU

 

 

 

 

CFG DAS RNGE PASS CLK MORE

 

EXIT

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

DIL only appears

SETUP C.3

 

RANGE CONTROL MENU

 

 

 

 

 

 

 

 

 

 

 

 

if the dilution ratio

 

 

 

 

 

 

 

 

 

option has been

MODE SET UNIT DIL

 

 

EXIT

 

 

installed

 

 

 

 

 

 

 

 

EXIT ignores the

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

new setting.

Toggle these buttons to set the

SETUP C.3

 

DIL FACTOR: 1.0 GAIN

 

 

ENTR accepts the

 

 

 

 

 

 

 

 

dilution factor.

0

0

0

1

.0

ENTR

EXIT

 

new setting.

 

 

 

This is the number by which the

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

analyzer will multiply the CO2

 

 

 

 

 

 

 

 

 

concentrations of the gas passing

SETUP C.3

 

DIL FACTOR: 20.0 GAIN

 

 

 

through the reaction cell.

 

 

 

 

 

 

 

 

 

 

0

0

2

0

.0

ENTR

EXIT

The analyzer multiplies the measured gas concentrations with this dilution factor and displays the result.

NOTE

Once the above settings have been entered, the instrument needs to be

recalibrated using one of the methods discussed in Section 7.