Image 105

Operating Instructions

Model GFC7000TA Carbon Dioxide Analyzer

6.7.2.12. HOLDOFF Feature

The DAS HOLDOFF feature allows to prevent data collection during calibrations and during the DAS_HOLDOFF period enabled and specified in the VARS (Section 6.12). To enable or disable the HOLDOFF, follow the instruction shown in Section 6.7.2.2 then press:

From the DATA ACQUISITION menu

(see Section 6.7.2.2)

Edit Data Channel Menu

SETUP X.X

0) CONC:

ATIMER,

1,

900

 

PREV NEXT

INS

DEL EDIT

PRNT

EXIT

 

 

 

 

 

 

 

 

 

 

 

 

 

SETUP X.X

NAME:CONC

 

 

 

<SET SET> EDIT PRINT

 

EXIT

 

 

 

 

 

 

 

Exits to the main Data Acquisition menu

 

 

Press SET> key until…

 

SETUP X.X

CAL HOLD OFF:ON

 

 

SET> EDIT PRINT

EXIT

Toggle to turn HOLDOFF

SETUP X.X

CAL HOLD OFF:ON

 

ON

 

ENTR EXIT

ON or OFF

 

ENTR accepts the new setting and returns to the previous menu.

EXIT ignores the new setting and returns to the previous menu.