Operating Instructions

Model GFC7000TA Carbon Dioxide Analyzer

FROM ANALOG I/O CONFIGURATION MENU

DIAG

 

ANALOG I / O CONFIGURATION

PREV

NEXT

ENTR EXIT

 

 

 

 

 

 

 

DIAG AIO

 

AIN CALIBRATED: NO

SET>

EDIT

EXIT

 

 

 

 

Press SET> to select the analog output channel

to be configured:.

DIAG AIO CONC_OUT_2:CURR, NO CAL

< SET SET> EDIT

EXIT

DIAG AIO CONC_OUT_2 RANGE: CURR

 

<SET SET>

EDIT

EXIT

DIAG AIO CONC_OUT_2 CALIBRATED: NO

 

< SET

CAL

EXIT

DIAG AIO AUTO CALIBRATING CONC_OUT_2

The instrument attempt to automatically calibrate

the channel … then beep.

DIAG AIO CONC_OUT_2 D/A/ CAL ERROR

EXIT

DIAG AIO CONC_OUT_2 CURR-Z: 0 mV

U100 UP10 UP DOWN DN10 D100 ENTR EXIT

DIAG AIO CONC_OUT_2 ZERO: 27 mV

U100 UP10 UP DOWN DN10 D100 ENTR EXIT

DIAG AIO CONC_OUT_2 SPAN: 10000 mV

U100 UP10 UP DOWN DN10 D100 ENTR EXIT

DIAG AIO CONC_OUT_2 ZERO: 9731 mV

U100 UP10 UP DOWN DN10 D100 ENTR EXIT

DIAG AIO CONC_OUT_2 CALIBRATED: YES

< SET

CAL

EXIT

Increase or decrease the current output by 100, 10 or 1 counts. The resulting change in output voltage is displayed in the upper line.

Continue adjustments until the correct current is measured with the current meter.

EXIT ignores the

new setting, ENTR

accepts the new

setting.

If a current meter is not available, an alternative method for calibrating the current loop outputs is to connect a 250 Ω ± 1% resistor across the current loop output. Using a voltmeter connected across the resistor, follow the procedure above but adjust the output to the following values:

Table 6-21: Current Loop Output Calibration with Resistor

Full scale

Voltage for 2-20 mA

Voltage for 4-20 mA

(measured across resistor)

(measured across resistor)

 

0%

0.5 V

1.0 V

100%

5.0 V

5.0 V

Teledyne Analytical Instruments

133

Page 151
Image 151
Teledyne gfc 7000ta operation manual Current Loop Output Calibration with Resistor, Set

gfc 7000ta specifications

The Teledyne GFC 7000TA is a cutting-edge, high-performance gas flow calibrator designed for both laboratory and field applications. This instrument is renowned for its accuracy, versatility, and reliability, making it an ideal choice for various industries, including aerospace, environmental monitoring, and manufacturing.

One of the standout features of the GFC 7000TA is its advanced measurement technology. It utilizes state-of-the-art thermal mass flow meters, which allow for precise fluid measurement across a wide range of flow rates. This technology ensures highly accurate readings, which are essential for applications requiring stringent compliance with regulatory standards.

The GFC 7000TA is capable of accommodating various types of gases, making it particularly versatile. Users can calibrate flow for gases such as air, nitrogen, or carbon dioxide, providing flexibility in a multitude of applications. The device also supports dual-channel capabilities, allowing simultaneous measurement and calibration of two different gas types, enhancing productivity and efficiency in processes.

Another notable characteristic of the GFC 7000TA is its user-friendly interface. Equipped with a high-resolution touchscreen display, the device offers intuitive navigation and data visualization. Users can easily set parameters, view real-time measurements, and generate printed reports. This interface reduces the learning curve for new operators, allowing organizations to implement the GFC 7000TA with minimal training.

In addition to its advanced measuring capabilities, the GFC 7000TA features a robust design for durability in harsh environments. The instrument is compliant with various international standards, ensuring its reliability and performance in various conditions. Its portable form factor makes it convenient for fieldwork, while the rugged construction provides peace of mind for users working in demanding settings.

The GFC 7000TA also incorporates modern connectivity options. With built-in USB and Ethernet ports, users can easily integrate the device into their existing systems for data logging and remote monitoring. This connectivity feature enables real-time data access and enhances the potential for analysis and optimization of gas flow systems.

In summary, the Teledyne GFC 7000TA is a high-performance gas flow calibrator characterized by its advanced thermal mass flow measurement technology, versatility in accommodating various gases, user-friendly interface, and robust design. With its modern connectivity features, this instrument stands out as a vital tool for professionals seeking precision and reliability in gas flow applications.