Teledyne gfc 7000ta operation manual Setup  CFG Viewing the Analyzer’s Configuration Information

Models: gfc 7000ta

1 280
Download 280 pages 4.46 Kb
Page 88
Image 88

Operating Instructions

Model GFC7000TA Carbon Dioxide Analyzer

6.5.SETUP CFG: Viewing the Analyzer’s Configuration Information

Pressing the CFG button displays the instrument configuration information. This display lists the analyzer model, serial number, firmware revision, software library revision, CPU type and other information. Use this information to identify the software and hardware when contacting customer service. Special instrument or software features or installed options may also be listed here.

Press NEXT and PREV to scroll through the following list of Configuration information:

MODEL NAME

PART NUMBER

SERIAL NUMBER

SOFTWARE REVISION

LIBRARY REVISION

iCHIP SOFTWARE REVISION1

HESSEN PROTOCOL REVISION1

ACTIVE SPECIAL SOFTWARE OPTIONS1

CPU TYPE

DATE FACTORY CONFIGURATION SAVED

 

SAMPLE*

RANGE = 500.000 PPB

CO2 =X.XXX

 

 

 

< TST TST > CAL

SETUP

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

SAMPLE

PRIMARY SETUP MENU

 

 

Press EXIT at any

 

 

 

 

 

 

 

time to return to the

 

CFG DAS RNGE PASS CLK MORE

EXIT

 

SAMPLE display

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

SAMPLE

T360 CO2 ANALYZER

 

 

Press EXIT at

 

 

 

 

 

 

 

any time to

 

NEXT PREV

 

EXIT

 

return to SETUP

 

 

 

 

 

 

 

 

 

 

menu

1Only appears if relevant option of Feature is active.

6.6. SETUP ACAL: Automatic Calibration

Instruments with one of the internal valve options installed can be set to automatically run calibration procedures and calibration checks. These automatic procedures are programmed using the submenus and functions found under the ACAL menu.

A menu tree showing the ACAL menu’s entire structure can be found in Appendix A-1 of this manual.

Instructions for using the ACAL feature are located in the Section 7.6 of this manual along with all other information related to calibrating the analyzer.

Teledyne Analytical Instruments

70

Page 88
Image 88
Teledyne gfc 7000ta Setup  CFG Viewing the Analyzer’s Configuration Information, Setup  Acal Automatic Calibration