Teledyne gfc 7000ta operation manual Default DAS Channels Setup

Models: gfc 7000ta

1 280
Download 280 pages 4.46 Kb
Page 93
Image 93

Operating Instructions

Model GFC7000TA Carbon Dioxide Analyzer

TEMP: Samples the analyzer’s bench temperature, box temperature and PHT cooler drive voltage every five minutes and records an average once every six hours. Data from the last 400 averaging periods is recorded. A time and date stamp is recorded for every data point logged. This information is useful for performing predictive diagnostics as part of a regular maintenance schedule (See Section 9.1).

Note

The CALDAT, STBZRO and STBSPN channels collect data based on events (e.g. a calibration operation) rather than a timed interval. This does not represent any specific length of time since it is dependent on how often calibrations are performed.

Triggering Events and Data Parameters/Functions for these default channels are:

LIST OF PARAMETERS

LIST OF CHANNELS

LIST OF PARAMETERS

PARAMETER: SMPLFLW MODE: AVG

NAME: PNUMTC

NAME: CONC

EVENT: ATIMER

REPORT PERIOD: 000:01:00

NO. OF RECORDS: 800

RS-232 REPORT: OFF

CHANNEL ENABLED: ON

CAL HOLD OFF: ON

PARAMETER: CONC1

MODE: AVG

PRECISION: 1

STORE NUM SAMPLES OFF

PRECISION: 1

STORE NUM SAMPLES OFF

PARAMETER: SMPLPRS

MODE: AVG

PRECISION: 1

STORE NUM SAMPLES OFF

PARAMETER: STABIL

MODE: INST

PRECISION:2

STORE NUM SAMPLES OFF

PARAMETER: DETMES MODE: INST

PRECISION: 1

STORE NUM SAMPLES OFF

PARAMETER: RATIO

MODE: INST

PRECISION: 3

STORE NUM SAMPLES OFF

PARAMETER: BNTEMP

MODE: AVG

PRECISION:1

STORE NUM SAMPLES OFF

PARAMETER: BOXTMP MODE: AVG

PRECISION: 1

STORE NUM SAMPLES OFF

PARAMETER: PHTDRV

MODE: AVG

PRECISION: 1

STORE NUM SAMPLES OFF

EVENT: ATIMER

REPORT PERIOD: 001:00:00

NO. OF RECORDS: 360

RS-232 REPORT: OFF

CHANNEL ENABLED: ON

CAL HOLD OFF: OFF

NAME: CALDAT

EVENT: SLPCHG

REPORT PERIOD: N/A

NO. OF RECORDS:200

RS-232 REPORT: OFF

CHANNEL ENABLED: ON

CAL HOLD OFF: OFF

NAME: STBZRO

EVENT: EXITZR

REPORT PERIOD: N/A

NO. OF RECORDS:200

RS-232 REPORT: OFF

CHANNEL ENABLED: ON

CAL HOLD OFF: OFF

NAME: STBSPN

EVENT: EXITSP

REPORT PERIOD: N/A

NO. OF RECORDS:200

RS-232 REPORT: OFF

CHANNEL ENABLED: ON

CAL HOLD OFF: OFF

NAME: TEMP

EVENT: ATIMER

REPORT PERIOD: 000:06:00

NO. OF RECORDS:400

RS-232 REPORT: OFF

CHANNEL ENABLED: ON

CAL HOLD OFF: OFF

PARAMETER: SLOPE1

MODE: INST

PRECISION:3

STORE NUM SAMPLES OFF

PARAMETER: OFSET1

MODE: INST

PRECISION: 1

STORE NUM SAMPLES OFF

PARAMETER: ZSCNC1

MODE: INST

PRECISION: 1

STORE NUM SAMPLES OFF

PARAMETER: DETMES

MODE: INST

PRECISION: 1

STORE NUM SAMPLES OFF

PARAMETER: RATIO

MODE: INST

PRECISION: 3

STORE NUM SAMPLES OFF

Figure 6-4: Default DAS Channels Setup

Teledyne Analytical Instruments

75

Page 93
Image 93
Teledyne gfc 7000ta operation manual Default DAS Channels Setup