6.13.SETUP MORE DIAG: Using the Diagnostics Functions

A series of diagnostic tools is grouped together under the SETUPMOREDIAG menu. As these parameters are dependent on firmware revision (see Menu Tree A-6 in Appendix A). The individual parameters, however, are explained in more detail in the indicated in 6-15. These tools can be used in a variety of troubleshooting and diagnostic procedures and are referred to in many places of the maintenance and trouble-shooting s.

Table 6-15: GFC 7000TA Diagnostic (DIAG) Functions

DIAGNOSTIC FUNCTION AND MEANING

Front Panel

 

Mode Indicator

 

 

 

SIGNAL I/O: Allows observation of all digital and analog signals in

 

 

the instrument. Allows certain digital signals such as valves and

DIAG I/O

6.13.2

heaters to be toggled ON and OFF.

 

 

 

 

 

ANALOG OUTPUT: When entered, the analyzer performs an analog

 

 

output step test. This can be used to calibrate a chart recorder or

DIAG AOUT

6.13.3

to test the analog output accuracy.

 

 

 

 

 

ANALOG I/O CONFIGURATION: the signal levels of the instruments

 

 

analog outputs may be calibrated (either individually or as a

DIAG AIO

6.13.4

group). Various electronic parameters such as signal span, and

 

 

offset are available for viewing and configuration.

 

 

 

 

 

ELECTRIC TEST: The analyzer is performing an electric test. This

DIAG OPTIC

6.13.5

test simulates IR detector signal in a known manner so that the

proper functioning of the sync/demod board can be verified.

 

 

 

 

 

DARK CALIBRATION: The analyzer is performing a dark

 

 

calibration procedure. This procedure measures and stores the

DIAG ELEC

6.13.6

inherent dc offset of the sync/demod board electronics.

 

 

 

 

 

PRESSURE CALIBRATION: The analyzer records the current

 

 

output of the sample gas pressure sensor. This value is used by

DIAG PCAL

6.13.7

the CPU to compensate the CO2 concentration.

 

 

FLOW CALIBRATION: This function is used to calibrate the gas flow

DIAG FCAL

6.13.8

output signal. This setting is retained when exiting DIAG.

 

 

TEST CHAN OUTPUT: Configures the A4 analog output channel.

DIAG TCHN

6.13.9

 

 

 

Teledyne Analytical Instruments

120

Page 138
Image 138
Teledyne gfc 7000ta Setup  More  Diag Using the Diagnostics Functions, GFC 7000TA Diagnostic Diag Functions