6.15.4.3. Activating Hessen Protocol

The first step in configuring the GFC 7000TA to operate over a Hessen protocol network is to activate the Hessen mode for com ports and configure the communication parameters for the port(s) appropriately. Press:

Repeat the entire process to set up the COM2 port

SAMPLE*

RANGE = 500.000 PPM

CO2 =X.XXX

< TST TST > CAL

SETUP

 

 

 

 

 

SETUP X.X PRIMARY SETUP MENU

 

CFG DAS RNGE PASS CLK MORE

EXIT

 

 

 

SETUP X.X SECONDARY SETUP MENU

SETUP X.X COM1 QUIET MODE: OFF

NEXT

OFF

ENTR

EXIT

 

Continue pressing next until …

 

SETUP X.X COM1 HESSEN PROTOCOL : OFF

 

PREV NEXT

OFF

ENTR

EXIT

 

COMM VARS DIAG

ALRM

EXIT

 

 

 

 

 

 

 

 

Select which COMM

SETUP X.X

COMMUNICATIONS MENU

 

port to configure

 

 

 

 

 

ID INET

COM1

COM2

EXIT

The sum of the mode

 

 

 

 

 

 

 

 

IDs of the selected

SETUP X.X

COM1 MODE:0

 

modes is displayed here

 

 

 

 

 

SET>

EDIT

 

EXIT

 

 

 

 

 

SETUP X.X COM1 HESSEN PROTOCOL : ON

 

 

PREV NEXT

ON

ENTR

EXIT

 

 

 

 

 

 

 

 

 

SETUP X.X COM1 E,7,1 MODE: OFF

 

 

 

PREV NEXT

OFF

ENTR

EXIT

 

 

 

 

 

 

 

 

 

 

SETUP X.X COM1 E,7,1 MODE: ON

 

 

 

PREV NEXT

ON

ENTR

EXIT

 

 

 

 

 

 

Toggle OFF/ON to

change

activate/deactivate

selected mode.

ENTR accepts the new

settings

EXIT ignores the new

settings

6.15.4.4. Selecting a Hessen Protocol Type

Currently there are two versions of Hessen Protocol in use. The original implementation, referred to as TYPE 1, and a more recently released version, TYPE 2 that has more flexibility when operating with instruments that can measure more than one type of gas. For more specific information about the difference between TYPE 1and TYPE 2 download the Manual Addendum for Hessen Protocol from the Teledyne Instruments web site: http://www.teledyne-api.com/manuals/index.asp .

To select a Hessen Protocol Type press: