Teledyne gfc 7000ta operation manual Software Operation, Adaptive Filter, Windows CE, API Firmware

Models: gfc 7000ta

1 280
Download 280 pages 4.46 Kb
Page 231
Image 231

Theory of Operation

Model GFC7000TA Carbon Dioxide Analyzer

9.5. Software Operation

The Model GFC 7000TA Gas Filter Correlation Carbon Dioxide Analyzer has a high performance, VortexX86-based microcomputer running Windows CE. Inside Windows CE, special software developed by Teledyne Instruments interprets user commands via the various interfaces, performs procedures and tasks, stores data in the CPU’s various memory devices and calculates the concentration of the sample gas.

Windows CE

 

 

API FIRMWARE

 

 

Memory Handling

Analyzer Operations

 

Calibration Procedures

PC/104 BUS

DAS Records

Configuration Procedures

Calibration Data

Autonomic Systems

 

System Status Data

 

Diagnostic Routines

 

 

 

 

 

ANALYZER

 

 

HARDWARE

 

Interface Handling

 

Measurement

Sensor input Data

 

Touchscreen/Display

 

Algorithm

 

Analog Output Data

 

 

RS232 & RS485

PC/104 BUS

 

External Digital I/O

 

Linearization Table

 

 

Figure 9-18: Basic Software Operation

9.5.1. Adaptive Filter

The GFC 7000TA software processes the CO2 MEAS and CO2 REF signals, after they are digitized by the motherboard, through an adaptive filter built into the software. Unlike other analyzers that average the output signal over a fixed time period, the GFC 7000TA averages over a set number of samples, where each sample is 0.2 seconds. This is technique is known as boxcar averaging. During operation, the software automatically switches between two different length filters based on the conditions at hand. Once triggered, the short filter remains engaged for a fixed time period to prevent chattering.

During conditions of constant or nearly constant concentration the software, by default, computes an average of the last 750 samples, or approximately 150 seconds. This provides the calculation portion of the software with smooth stable readings. If a rapid change in concentration is detected the filter

Teledyne Analytical Instruments

213

Page 231
Image 231
Teledyne gfc 7000ta operation manual Software Operation, Adaptive Filter, Windows CE, API Firmware, Interface Handling