Agilent Technologies 66111A, D, 66311B, 66309B manual Quick Reference, Front Panel At a Glance

Page 15

1

Quick Reference

The Front Panel - At a Glance

1 A 14-character display

 

2 Annunciators indicate

3 Rotary control sets voltage,

shows output measurements

operating modes and status

current, and menu parameters.

and programmed values.

 

conditions.

Use ￿and «

 

 

 

 

 

to set the resolution; then adjust

 

 

 

the value with the knob.

1

2

 

3

 

66309D

DUAL OUTPUT

 

Mobile Communications DC Source

CV CC

Unr Dis

OCP Prot Cal Shift Rmt Addr Err SRQ

 

 

 

 

SYSTEM

 

 

FUNCTION

 

 

ENTRY

Channel

Error

 

Input

OV

 

Res

Cir Entry

Local

Address

 

Meter

Voltage

.

Output

Enter

LINE

 

 

 

 

 

Number

1

2

3

4

5

 

-

 

 

 

 

Save

 

Prot Cir

OCP

 

Cal

 

Off

Recall

 

Protect

Current

 

Output

Enter

On

6

7

8

9

0

On/Off

 

 

 

 

Backspace

4

 

5

6

 

 

 

 

 

4 Turns the dc

5 System keys:

6 Function keys:

source on and off.

return to Local mode

enable/disable the

 

select output channel

 

output

 

set GPIB address

select metering

 

set RS-232 interface

 

functions

 

display SCPI error

program voltage and

 

 

codes

 

current

 

save and recall

set and clear protection

 

 

instrument states

 

functions

 

display firmware

 

Û

and

Ä

 

 

 

 

 

 

revision and serial

 

 

 

 

 

 

 

 

 

scroll

through the front

 

 

number.

 

panel menu commands.

7

 

 

 

 

 

 

7 Entry keys:

 

 

 

enter values

 

 

 

increment or

 

 

 

decrement values

and

 

 

»

 

 

 

 

 

 

 

 

front

 

 

 

select

panel

 

 

menu parameters.

￿

 

«

 

 

and

 

 

 

select a digit in the numeric entry field.

15

Image 15
Contents USER’S Guide Certification Warranty InformationGeneral Safety SummarySafety Symbols EMC DeclarationDeclaration of Conformity Printing History Acoustic Noise InformationTable of Contents DVM Connections Checkout Procedure Case of TroubleVXIplug&play Power Products Instrument Drivers Types of Scpi CommandsTriggering Output Changes Types of Scpi MessagesScpi Data Formats Introduction Programming the OutputMeasurement Commands 104 Inhibit/Fault IndicatorCalibration Commands 100 Display Commands 103Trigger Commands 126 Output Commands 114Status Commands 121 System Commands 125Performing the Calibration Procedure 154 Common CommandsAdditional Commands Specifications 143 Supplemental Characteristics 144Introduction 177 Error Number List 161Introduction 165 BasicFront Panel At a Glance Quick ReferenceInstrument Configuration Use the front panel Address key to configure the interfaceRear Panel At a Glance Use the Function keys and Entry keys to enter a new value Front Panel Number EntryImmediate Action Keys Front Panel AnnunciatorsFront Panel Menus At a Glance ABORt CALibrate Scpi Programming Commands At a GlanceInstalling the VXIplug&play instrument driver Using the front panelUsing the programming interface Programming the unit using Scpi COMPatibility commandsDescription Safety ConsiderationsOptions and Accessories OptionAgilent 66309D Description and Model DifferencesAgilent Agilent 66309BCommon Capabilities Front Panel ControlsRemote Programming Dc Source Output 1 Characteristic Output 1 CharacteristicOutput 2 Characteristic Output 2 CharacteristicOption 521 Factory Settings Option 521 Description Agilent 66309B/D onlyOption 521 Relay Modes Page Check the Phone Connections Installation and Operation ChecklistCheck the Operating Settings and Conditions Additional Agilent 66311/66309 Operating Settings ChecksDamage Packaging Material CleaningInspection LocationRack Mounting Connect the Power CordInput Connections Bench OperationCurrent Ratings Turn the unit off before connecting any wiresOutput Connections OutputRemote Sense Connections Remote Sense ConnectionsRemote Sense Connections with External Relays Maintaining Stability while Remote Sensing Load Regulation and Voltage Drop in the Remote Sense LeadsOpen Sense Lead Protection Message Description Local SensingOutput Compensation OVP Considerations DVM ConnectionsMeasuring Circuits Not Powered by the Main Output Measuring Circuits that are Not Powered by the Main OutputMeasuring Circuits Floating with Respect to the Main Output 10. FLT/INH Examples External Protection ConnectionsGpib Interface FLT/INH DIGital I/O ConnectorDigital I/O Connections Computer ConnectionsPin Input/Output Description RS-232 InterfacePress Protect Checkout ProcedureProcedure Display Explanation Press Output On/OffProcedure Display Explanation Enter Number 12, Enter Press Output On/Off Selftest Error MessagesCase of Trouble Press Shift, ChannelLine Fuse Runtime Error MessagesPower-On Selftest Errors Runtime Error MessagesFront Panel Description IntroductionFront Panel Operation Display Command Function System KeysOFF Function Keys Immediate Action KeysScrolling Keysq Display Measurement Metering KeysOver Current Output Control KeysEntry Keys Entry KeysSelecting the DVM on Agilent 66311D/66309D units Using the Front Panel DisplayExamples of Front Panel Programming Selecting an output on Agilent 66309B/D unitsTypecap High Set the output voltageSet the output current limit Set the output compensationKeypad, press Enter Number, 7, Enter Enable the outputSet the output 2 voltage Set the output 2 current limitOvercurrent Querying and Clearing Output Protection and ErrorsDisable Overvoltage Protection as follows Query and Clear Errors as followsUse the Meter menu for making front panel measurements Making Basic Front Panel MeasurementsMaking Enhanced Front Panel Measurements Default Front Panel Measurement ParametersCurrdet Acdc Making DVM Measurements Agilent 66311D/66309D onlyUse the Meter menu for making DVM measurements Currrang AutoSet the Gpib address as follows Setting the Gpib Address and Programming LanguageTo configure the Ridfi mode of the port, proceed as follows To configure the Digio mode of the port, proceed as followsStoring and Recalling Instrument States External References Gpib ReferencesScpi References System Requirements VXIplug&play Power Products Instrument DriversDownloading and Installing the Driver Supported ApplicationsGpib Address Accessing Online HelpGpib Capabilities of the DC Source RS-232 Capabilities of the DC SourceIntroduction to Scpi RS-232 Flow ControlBaud Rate Types of Scpi Commands Conventions Used in This GuideBoldface font Multiple Commands in a Message Including Common CommandsMoving Among Subsystems Headers Using QueriesTypes of Scpi Messages Message UnitScpi Data Formats Class Suffix Unit Unit with Multiplier Scpi Command CompletionSuffixes and Multipliers Response Data TypesScpi Conformance Information Using Device ClearScpi Conformed Commands Non-SCPI CommandsPower-on Initialization Programming the OutputEnabling the Output Output Current Output VoltageOutput Trigger Model Setting the Voltage or Current Transient LevelsTriggering Output Changes Scpi Triggering NomenclatureSingle Trigger Enabling the Output Trigger SystemSelecting the Output Trigger Source Generating TriggersMaking Basic Measurements Average MeasurementsControlling Measurement Samples Window Functions Making Enhanced MeasurementsRMS Measurements Current Ranges and Measurement DetectorPulse Measurements Minimum and Maximum MeasurementsHigh/Low Measurements Returning All Measurement Data From the Data Buffer Making DVM MeasurementsTriggered Measurements Measurement Trigger ModelSequence Form Alias SEQuence2 ACQuire INTernal Enabling the Measurement Trigger SystemSelecting the Measurement Trigger Source Selecting the Sensing FunctionSingle Triggers Generating Measurement TriggersTrigacqcouncurr number or Trigacqcounvolt number Pre-trigger and Post-trigger Data Acquisition Programming the Status RegistersDC Source Status Model Power-On ConditionsOperation Status Group Bit Configurations of Status RegistersStatus Byte Register PON Power On BitQuestionable Status Group Standard Event Status GroupRQS Bit Determining the Cause of a Service InterruptServicing Operation Status and Questionable Status Events MSS BitInhibit/Fault Indicator Monitoring Both Phases of a Status TransitionRemote Inhibit RI Pin Discrete Fault Indicator DFIUsing the Inhibit/Fault Port as a Digital I/O Bit WeightSubsystem Commands Subsystem Commands Syntax Language Dictionary Language Dictionary Programming Parameters Common CommandsCALibrateCURRentMEASureLOWRange Calibration CommandsCALibrateCURRent CALibrateCURRent2CALibrateDVM CALibratePASSwordCALibrateDATA CALibrateDATECALibrateVOLTage2 CALibrateSAVECALibrateSTATe CALibrateVOLTageDISPlayMODE Display CommandsDISPlay DISPlayCHANnelMeasurement Commands FORMat104 Query Syntax FORMatBORDerMEASureARRayCURRent? FETChARRayCURRent? MEASureARRayVOLTage? FETChARRayVOLTage?NR3 MEASureCURRent? FETChCURRent?MEASureCURRent2? MEASureCURRentACDC? FETChCURRentACDC?107 MEASureCURRentHIGH? FETChCURRentHIGH?MEASureCURRentLOW? FETChCURRentLOW? MEASureCURRentMAXimum? FETChCURRent MAXimum?MEASureVOLTage? FETChVOLTage? MEASureCURRentMINimum? FETChCURRentMINimum?MEASureDVM? FETChDVM? MEASureDVMACDC? FETChDVMACDC?109 MEASureVOLTage2MEASureVOLTageACDC? FETChVOLTageACDC? MEASureVOLTageHIGH? FETChVOLTageHIGH?110 MEASureVOLTageLOW? FETChVOLTageLOW?MEASureVOLTageMAXimum? FETChVOLTageMAXimum? MEASureVOLTageMINimum? FETChVOLTageMINimum?111 SENSeCURRentDETectorSENSeCURRentRANGe Query Syntax SENSeCURRentDETector?SENSeSWEepPOINts SENSeFUNCtionSENSePROTectionSTATe SENSeSWEepOFFSetPOINtsQuery Syntax SENSeWINDowTYPE? SENSeSWEepTINTervalSENSeWINDow Query Syntax SENSeSWEepTINTerval?OUTPut1 2RELayMODE Output CommandsINSTrumentCOUPleOUTPutSTATe OUTPut1115 OUTPutDFIOUTPutDFISOURce OUTPutPONSTATeOUTPutTYPE OUTPutPROTectionDELayOUTPutPROTectionCLEar OUTPutRIMODESOURceCURRent SOURceCURRent2117 SOURceCURRent2TRIGger Command Syntax Parameters RST Value ExamplesSOURceCURRentPROTectionSTATe SOURceCURRentTRIGger119 SOURceDIGitalDATASOURceDIGitalFUNCtion SOURceVOLTage120 SOURceVOLTage2SOURceVOLTagePROTection SOURceVOLTagePROTectionSTATeSOURceVOLTage2TRIGger Status CommandsSTATusPRESet SOURceVOLTageTRIGgerSTATusOPERationCONDition? STATusOPERationENABleBit Configuration of Operation Status Registers STATusOPERation?STATusQUEStionable? Parameters Preset ValueBit Configuration of Questionable Status Registers STATusOPERationNTR STATusOPERationPTR124 STATusQUEStionableENABleSTATusQUEStionableCONDition? STATusQUEStionableNTR STATusQUEStionablePTRSYSTemVERSion? System CommandsSYSTemERRor? SYSTemLANGuageINITiateSEQuence INITiateNAME Trigger CommandsRelated Commands ABORtTRIGger TRIGgerSOURce127 128 TRIGgerSEQuence2 TRIGgerACQuireTRIGgerSEQuence2COUNtCURRent TRIGgerACQuireCOUNtCURRent TRIGgerSEQuence2COUNtDVM TRIGgerACQuireCOUNtDVMTRIGgerSEQuence2COUNtVOLTage TRIGgerACQuireCOUNtVOLTage Unit129 130 TRIGgerSEQuence2HYSTeresisDVM TRIGgerACQuireHYSTeresisDVMTRIGgerSEQuence2LEVelCURRent TRIGgerACQuireLEVelCURRent TRIGgerSEQuence2LEVelDVM TRIGgerACQuireLEVelDVM131 132 TRIGgerSEQuence2LEVelVOLTage TRIGgerACQuireLEVelVOLTageTRIGgerSEQuence2SLOPeCURRent TRIGgerACQuireSLOPeCURRent Returned ParametersTRIGgerSEQuence2SLOPeDVM TRIGgerACQuireSLOPeDVM TRIGgerSEQuence2SLOPeVOLTage TRIGgerACQuireSLOPeVOLTage133 TRIGgerSEQuence2SOURce TRIGgerACQuireSOURce TRIGgerSEQuence1DEFine TRIGgerSEQuence2DEFine134 135 Common CommandsCommand Syntax *CLS Parameters None Bit Configuration of Standard Event Status Enable RegisterField Information Example136 137 Query Syntax *OPT? Returned Parameters AardRCL NRf Parameters ExampleCommand Syntax *SAV NRf Example *SAV Related Commands *RCL *RST 138RST Settings 139 Power-on ValueBit Configuration of Status Byte Register Query Syntax *STB?Query Syntax TST? Returned Parameters NR1 TRGAdditional Commands VOLTagePROTectionTRIPped? CURRentPROTectionTRIPped?VOLTageLIMitHIGH? VOLTageLIMitLOW?Specifications Table A-1. Performance Specifications143 Agilent 66309B/D Supplemental CharacteristicsTable A-2. Supplemental Characteristics 144145 146 Table A-4. Agilent 66309B/D Option 521 CharacteristicsTable B-1. Equipment Required Test SetupEquipment Required 147Performing the Verification Tests Turn-On Checkout148 Steps 6-10 apply to Agilent 66309B/D output 2 only Voltage Programming and Measurement AccuracyCurrent Programming and Measurement Accuracy Output terminals on both outputs 1Steps 7-11 apply to Agilent 66309B/D output 2 only Current Sink Measurement150 151 DVM Measurement Accuracy152 DVM Voltage Measurement Agilent 66309D only 153Performing the Calibration Procedure Front Panel Calibration Menu154 Voltage Programming and Measurement Calibration Front Panel Calibration ProcedureCheck the Language Setting Enable Calibration Mode156 Steps 11-16 apply to Agilent 66309B/D output 2 onlyOvervoltage Protection Calibration Current Programming and High-Range Measurement CalibrationSteps 27-32 apply to Agilent 66309B/D output 2 only 157Calcurrmeas AC 158 Restore the Language SettingDVM Calibration applies to Agilent 66311D, 66309D only Saving the Calibration ConstantsCalibration Over the Gpib Calibration Error MessagesChanging the Calibration Password Table B-3. Gpib Calibration Error MessagesPage Error Number List Table C-1. Error Numbers161 162 Error Messages163 Error Messages CPage Assigning the Gpib Address in Programs National Instruments Gpib Driver165 Error Handling Example 1. National Instruments Interface Example166 Example Programs D 167168 Example 2. Controller Using Basic169 Example 3. Current Pulse Measurement Using Basic170 171 172 Example 5. DFI Example Using Basic DFI Programming Example173 Page Close the Unit Configure the Power TransformerInstall the Correct Line Fuse Open the Unit176 Figure E-1, Power Transformer AC Input ConnectionsCommand Setting Table F-1. COMPatibility Power-on Settings177 178 Compatibility CommandSimilar Scpi Command Table F-2. COMPatibility Commands179 180 ErrorNumberError String Description/Explanation/Examples Table F-3. COMPatibility Errors181 Table F-5. Bit Configuration of Serial Poll RegisterPage 183 Index184 Index185 186 OVERTEMPERATURE, 47 OVERVOLTAGE, 47 OVLD, 47, 60, 61187 Subsystem commands syntax, 96 suffixesType CAP 188Japan United States Latin AmericaCanada Australia/New Zealand Europe Asia PacificManual Updates
Related manuals
Manual 8 pages 56.17 Kb

66111A, 66309B, 66311B, D specifications

Agilent Technologies D,c,83440b is an advanced electronic measurement solution designed for engineers and scientists who require precise and reliable performance in their testing environments. This modular test system offers a comprehensive suite of features that cater to a wide range of applications, from high-frequency testing to complex signal analysis.

One of the main features of the D,c,83440b is its impressive frequency range, allowing users to conduct tests across a wide spectrum of signals. The system is capable of handling frequencies up to 26.5 GHz, making it ideal for RF and microwave applications. This broad range ensures that users can work with a variety of devices, including communication systems, radar, and satellite technology.

In addition to its frequency capabilities, Agilent Technologies has engineered the D,c,83440b with exceptional dynamic range and low noise figures. This ensures that even the smallest signals can be accurately measured, allowing for greater precision in testing. The full spectrum analysis feature enables users to capture transient events and analyze them in real-time, which is crucial for troubleshooting and performance evaluations.

The D,c,83440b is built on a modular platform, allowing users to customize their systems according to specific testing needs. This modularity not only enhances flexibility but also simplifies maintenance and upgrades. Users can easily swap out different modules without the need for extensive system reconfiguration, which can significantly reduce downtime in testing environments.

Another standout characteristic of the D,c,83440b is its user-friendly interface. With a large, high-resolution display and intuitive controls, engineers can quickly navigate through settings and data, streamlining the testing process. This ease of use is complemented by powerful software solutions that can automate test sequences, aiding in efficiency and accuracy.

The integration of advanced digital signal processing technologies further enhances the capabilities of the D,c,83440b. These technologies enable more sophisticated measurements and improved signal integrity, which is essential for modern communication systems.

In summary, the Agilent Technologies D,c,83440b is a multifaceted electronic measurement solution that boasts a wide frequency range, excellent dynamic range, modular design, and user-friendly interface. This combination of features makes it suitable for various applications, ensuring that engineers and scientists have the tools they need to succeed in their testing and measurement endeavors.