Figure 8-2 -- Even though the star pattern appears the same on both sides of focus, they are asymmetric. The dark obstruction is skewed off to the left side of the diffraction pattern indicating poor collimation.

To accomplish this, you need to tighten the secondary collimation screw(s) that move the star across the field toward the direction of the skewed light. These screws are located in the secondary mirror holder (see figure 8-1). To access the collimation screws you will need to rotate the collimation screw cover clockwise to expose the three collimation screws underneath. Make only small 1/6 to 1/8 adjustments to the collimation screws and re-center the star by moving the scope before making any improvements or before making further adjustments.

To make collimation a simple procedure, follow these easy steps:

1.While looking through a medium to high power eyepiece, de-focus a bright star until a ring pattern with a dark shadow appears (see figure 8-2). Center the de-focused star and notice in which direction the central shadow is skewed.

2.Place your finger along the edge of the front cell of the telescope (be careful not to touch the corrector plate), pointing towards the collimation screws. The shadow of your finger should be visible when looking into the eyepiece. Rotate your finger around the tube edge until its shadow is seen closest to the narrowest portion of the rings (i.e. the same direction in which the central shadow is skewed).

3.Locate the collimation screw closest to where your finger is positioned. This will be the collimation screw you will need to adjust first. (If your finger is positioned exactly between two of the collimation screws, then you will need to adjust the screw opposite where your finger is located).

4.Use the hand control buttons to move the de-focused star image to the edge of the field of view, in the same direction that the central obstruction of the star image is skewed.

5.While looking through the eyepiece, use an Allen wrench to turn the collimation screw you located in step 2 and 3. Usually a tenth of a turn is enough to notice a change in collimation. If the star image moves out of the field of view in

the direction that the central shadow is skewed, than you are turning the collimation screw the wrong way. Turn the screw in the opposite direction, so that the star image is moving towards the center of the field of view.

Figure 8-3

A collimated telescope

should appear

symmetrical with the

central obstruction centered in the star's diffraction pattern.

6.If while turning you notice that the screws get very loose, then simply tighten the other two screws by the same amount. Conversely, if the collimation screw gets too tight, then loosen the other two screws by the same amount.

7.Once the star image is in the center of the field of view, check to see if the rings are concentric. If the central obstruction is still skewed in the same direction, then

continue turning the screw(s) in the same direction. If you find that the ring pattern is skewed in a different direction, than simply repeat steps 2 through 6 as described above for the new direction.

Perfect collimation will yield a star image very symmetrical just inside and outside of focus. In addition, perfect collimation delivers the optimal optical performance specifications that your telescope is built to achieve.

If seeing (i.e., air steadiness) is turbulent, collimation is difficult to judge. Wait until a better night if it is turbulent or aim to a steadier part of the sky. A steadier part of the sky is judged by steady versus twinkling stars.

53

Image 53
Contents CGE800 CGE925 CGE1100 CGE1400 User Defined Objects Telescope Maintenance Some of the many standard features of the CGE include 12v Output Jack CGE 1400 ShownPower Supply Setting up the TripodCounterweight DiagonalAttaching the Electronics Pier Attaching the Center Leg BraceInstalling the Counterweight Installing the Counterweight BarCounterweight Locking Screw Safety Screw Washer Attaching the Equatorial MountDovetail Slide Bar Dovetail Locking Knobs Mounting Platform Attaching the Optical Tube to the MountAttaching the Visual Back Installing the Eyepiece Installing the Star DiagonalInstalling the Finderscope Moving the Telescope Manually Adjusting the MountBalancing The Mount in DEC Balancing The Mount in R.ATransporting the CGE Powering the TelescopeAttaching the Motor Cables Page CGE Hand Control Named Objects Named StarsDouble Stars Variable StarsStartup Procedure Alignment ProceduresAuto One-Star Align Auto Two-Star AlignCGE Re-Alignment Quick-AlignLast Alignment Selecting an Object Object CatalogSlewing to an Object Finding PlanetsConstellation Tour Tour ModeDirection Buttons Rate ButtonSetup Procedures Identify Precise GoTo Scope Setup FeaturesObserving Tip Helpful Hint Utility Features Helpful Hint Page CGE Ready Image Orientation Determining Field of View FocusingCalculating Magnification General Observing Hints Celestial Coordinate System Motion of the Stars Pointing at Polaris Latitude ScalesFinding the North Celestial Pole Declination Drift Method of Polar Alignment Lunar Observing Hints Observing the MoonObserving the Planets Planetary Observing HintsSolar Observing Hints Observing the SunObserving Deep Sky Objects Seeing ConditionsSeeing Short Exposure Prime Focus Photography Full Eyepiece ProjectionPlanet ISO Moon Mercury Venus Mars Jupiter Saturn Long Exposure Prime Focus PhotographyPeriodic Error Correction PEC Using Periodic Error Correction Metering Terrestrial PhotographyReducing Vibration CCD ImagingFastar Compatible Optical System Fastar F/2 Imaging Planetary or Lunar With Reducer/CorrectorMedium size to small galaxies Auto Guiding Collimation Care and Cleaning of the OpticsCollimated telescope Should appear Symmetrical with Page At f/10 and 163 power with the CGE Page Technical Specifications Idle Current Appendix a Longitudes Latitudes Georgia Minnesota Rhode Island Canada Communication Protocol Appendix B RS-232 ConnectionReset The Position Of Azm Or Alt Additional RS232 CommandsAppendix C Maps of Time Zones Page Page Page Page Page Page Page Celestron TWO Year Warranty