Tripp Lite HCRK-3 owner manual Power Supply Module Battery Module

Page 9

Features continued

Power Supply Module continued:

Battery Module:

USB Port: This port connects the Power Supply Module to any cart- mounted computer (laptop or desktop) with included cable. Use with PowerAlert Data-Saving Software (included on CD) to automatically save open patient data files and safely close applications if the Battery Module is depleted or if utility power is disconnected while the cart is unattended. Consult your PowerAlert manual (included on CD) for more information.

Battery Connectors: Use to connect the Power Supply Module to the Battery Module. Contact Tripp Lite for additional solutions to meet all your runtime needs.

Cooling Fan: Quiet, efficient fan regulates internal temperature and prolongs equipment service life. Fan runs intermittently depending on temperature and load.

Communications Jacks: Connects the Power Supply Module to the Remote User Interface (RUI).

Isolation Transformer (Internal. Not Shown): Ensures full

UL 60601-1 compliance by managing current leakage of connected equipment.

Automatic Transfer Switch—ATS (Internal. Not Shown): Works automatically, whether plugged in or mobile. Allows Power Supply Module to simultaneously run connected equipment while recharging the Battery Module.

Advanced, 3-Stage Battery Charger (Internal. Not Shown): Recharges batteries safer and faster than conventional chargers. Faster charging ensures faster cart rotation between shifts.

Battery Cable: Use to connect the Battery Module to the Power Supply Module.

DC Fuses: Protect Battery Module from damaging overload. If fuse blows, replace with standard 25 amp automotive fuse. CAUTION: installing non-rated fuses could cause equipment damage and void your warranty.

9

Image 9
Contents Medical-Grade Mobile Power Retrofit Kit Power Supply Module & Battery Module Location Warnings Important Safety InstructionsPower Supply Module Connection Warnings Equipment Connection WarningsLabeling Symbols Battery Module WarningsPower Supply Module & Battery Module Mounting MountingRemote User Interface RUI Mounting Plug In the Power Supply Module Connect the Battery Module to the Power Supply ModuleCharge the Battery Module Connection / Start-UpOPTIONAL-Install PowerAlert Data-Saving Software Turn the Power Supply Module’s Outlets onStationary Operation Plugged In / Battery Charging Mobile Operation Unplugged / Battery DischargingRemote User Interface RUI Power Supply Module Battery Module Power button after reconnection. See Connection / Start-Up Storage ServiceWarranty Registration Limited Warranty