Omega Speaker Systems DPS3301 manual Timers, Open Collector Option

Page 17

limit alarms. (Check model number printed on the unit for option). If ordered with relays, then these relays are programmed during Setup to operate as Normally Open (NO) or Normally Closed (NC). The default setting is Normally Open. Each of these limit relays provides a switched output whenever a limit is reached. The maximum rating for a 120Vac unit is 120Vac @ 0.5 amp or 28 Vdc @ 1.0 amp.

Figure 2. Electro-mechanical Relays Hookup Example

Limit# #4

Limitt# 3#

Instrument

Limit# 2#

Limit# 1#

8

7

6

5

4

3

2

1

PLUG-IN CONNECTOR

Switched

SWITCHED OUTPUT

 

 

120 VAC @ 0.5 AMPS

120 VAC @ 0.5 AMPS

120 VAC @ 0.5 AMPS

120 VAC @ 0.5 AMPS

Open Collector Option

Whenever a limit is reached (on units with open collector output option) an open collector output provides a return for 5 Vdc signal at 50ma on the output connector, e.g. Limit 1 provides its 5 volt output signal between pins 1 and 2. A solid state control relay with a 5Vdc coil that draws 50 ma. or less is recommended for use as shown in figure 4 below. The monitor supplies +5Vdc on pin #2 and the open collector switch to ground on pin #1. The control relay should be driven by controller’s internal power supply off of pin #2 to avoid introducing ground loops or electrical noise into the unit.

Figure 3. Open Collector Hookup Example

Instrument

2

1

+OUTPUT

-OUTPUT

LOAD

Solid State Relay, LED, Buzzer etc.

TIMERS

The system features two timers that can be programmed from 1 - 99999 seconds. These timers come in handy when some time- based functions have to be performed. For example, it is possible to start a vacuum pump after a certain amount of preprogrammed time or to shut down a process a certain amount of time after reaching a preprogrammed temperature ,etc. Timers can be programmed to operate in one of three possible modes (see below).

Page 17

Image 17
Contents DPS3301 Series Wall-MountProgrammable Temperature Monitor Specifications Description Thermocouple or RTD Selection Timers Accuracy SpecificationsThermocouple or RTD Selection DescriptionDisplay Modes How to Reset and Program Thermocouple Type or RTDDisplay Hold Feature Retaining User Settings on Power DownDisplay Units C/F Setpoint DeviationProgram Limits Flow ChartEntering and Exiting Setup Mode Setup ModeHow To Increase/Decrease Programmable Values Setup Passcode EntryDisplay Time Program Controller ModeDisplay Option Rate Time-BaseThermocouple Calibration Cold Junction adjustmentThermocouple type Selection ‘PROGRAM LIMITS’ ModeProcess Limits 1 thru Selection of Limit 3 as Process or Rate LimitSelection of Limit 4 as Process or Deviation Configure Timing ModeProgramming Limit Configuring Limits Normally Open/Normally Closed Timers 1Relay Deadband Relay Latching/non LatchingAudio Alarm On/Off Setpoint Deviation AlarmRate Alarm Resetting Rate AlarmProcess Alarms Open Collector Option TimersElapsed Time Timer ModesSignal Connection PowerHow to Reset High & Low Peak readings High and LOW PeaksOutput Connector Pin Assignment Output Connector PIN AssignmentDPS3301 Series Terminal Compartment Direct Interface to Universal Relay Module Mounting Page Page M4300/0907