Omega Vehicle Security CN3271 Series manual Logic Input for models with RS-485 only, Relay Outputs

Page 8

B) Logic Input (for models with RS-485 only)

This input is used for remote acknowledgement (reset).

Safety note:

-Do not run logic input wiring with AC power cables.

-Use an external dry contact capable of switching 0.5 mA, 5 Vdc.

-The instrument needs 100 ms to recognize a contact status variation.

-The logic inputs are NOT isolated from the measuring input.

Logic input

14

C.1) Relay Outputs

OUT 1 1

Class 1

2

 

3

OUT 2

6

Class 1

 

(Alarm)

 

 

7

Fig. 8

 

NO - OUT 1

C - OUT 1

NC - OUT 1

C

NO

15

Fig.7

The OUT 1 contact rating is 3A/250V AC on resistive load.

The OUT 2 contact rating is 2A/250V AC on resistive load.

The number of operations is 1 x 105 at specified rating.

NOTES 1) To avoid electric shock, connect power line at the end of the wiring procedure.

2)For power connections use No 16 AWG or larger wires rated for at last 75 °C.

3)Use cupper conductors only.

4)Don’t run input wires with power cables. All relay contacts are protected by varistor against inductive load with inductive component up to 0.5 A.

The following recommendations avoid serious problems which may occur, when relay outputs are used with inductive loads.

5

LHL-1-NEUTRO.pmd

5

19/04/2004, 15.07

Image 8
Contents CN3271-add2.pmd 19/08/2004 OMEGAnet On-LineService Internet e-mail Info@omega.com Contents Screw Bracket Panel Mounting RequirementsDimensions and Rear Terminal Blocks Without RS-485 With RS-485Panel Cutout Wiring Guidelines MA, mV and V Inputs Wiring RTD Input Linear InputLogic Input for models with RS-485 only Relay OutputsOUT 1 Class Use cupper conductors onlyInductive Loads Serial InterfacePower Line and grounding AA/A100 to 240 Vac 500 mA 250 100/240 V AC 125 mAHigh Low Limiter with RS-485 High Low Limiter without RS-485Serial Interface Protocol Configuration KEY FunctionsConfiguration Procedure R5 = Offset Adjustment R7 = Alarm action on input faultType Value R2 = Decimal Point PositionC2 = Rearming Mode Hi.Lo = High and low limit for special process= Rearm at Power-up Auto = Automatic rearm Constant Alarm Function= Alarm configuration = Shutdown memoryExample for P2 = H.L P3 = Alarm ActionExample for P2 = H.A.A.c T1 = Timeout Selection Alarm Standby mask Function= Safety Lock Operating Mode Normal Display ModeIndicators Flashes when alarm is onNormal Display Mode to Operative Parameter Display Mode Operative Parameter Display ModeOperating Parameters Limiter function Example 1 C1 = Hi and C2 = O Example 2 C1 = Hi and C2 =Device can be connected to a host computer via serial link Alarm functionsSerial LInk optional Error Messages Error MessagesError List Dealing with Error MessagesGeneral Specifications Power consumption 8 VACommon mode rejection ratio Temperature Drift CJ excludedInputs Thermocouple Linear InputsAlarm Logic InputsOutputs OutputMaintenance Serial Communication Interface OptionDefault Parameters Default Operating Parameters ListLHL-A-NEUTRO.pmd 19/04/2004 Second Seconds Appendix A.3 Appendix B Thermocouple Compensating Cable Color Codes Return Requests / Inquiries Data Acquisition