UltraViolet Devices V-Flex Specifications, Power Ratings, Temperature Ratings, Power Consumption

Page 3

SPECIFICATIONS

Electrical Ratings

Power Ratings

Note: The V-Flex is a universal input circuit capable of operation at 120 Vac or 240 Vac and 50Hz or 60Hz.

The unit is designed to be hardwired to an appropriate electrical service in accordance with NEC.

Temperature Ratings

Ambient Temperature Range: 40°F to 135°F (4°C to 51°C)

Lamp Temperature Range: 40°F to 135°F (4°C to 51°C)

Relative Humidity: Up to 95% rh, non-condensing

Power Consumption

Model

Input Current

Lamp Wattage

 

(A)

(W)

 

 

 

120V

0.6

60

 

 

 

240V

0.3

60

 

 

 

Approvals

Underwriters Laboratories: File no. E212213. Category Code ABQK (Accessories, Air Duct

Mounted) UL Standards: UL153, UL1598 & UL1995. The health aspects associated with the use of this product and its ability to aid in disinfection of environmental air have not been investigated by UL. This product is classified only when installed with accompanying part numbers 19-1020,19-1030,19-3004 and 19- 3005.

Figure 1. V-FlexTMLamp Module

19-2021 Rev E, 06/05/07

3

Image 3
Contents Flex Table of Contents Power Consumption SpecificationsPower Ratings Temperature RatingsInstallation ApplicationProduct Registration Mounting Location Dimension a Dimension B Dimension CInstallation of the Lamp Modules Rev E, 06/05/07 Installation of the Power Junction Module Preparation of the WIRE/RACEWAY CoverRaceway Installation of Main Support Into AIR Handling Duct Electrical Connection of the System Lamp InstallationRev E, 06/05/07 Rev E, 06/05/07 Operation Troubleshooting and Service Broken Lamp Clean-upMercury Notice Replacing the Lamp Annually Cleaning the LampsRev E, 06/05/07 Equipment Warranty Pureuvc661 257-4698 fax