Pressure Systems user manual Pressure Systems, Inc Model 9046 User’s Manual

Page 50

Pressure Systems, Inc.

Model 9046 User’s Manual

 

 

5.7.1 Command ’c’- Sub-command Index 00: Configure a Host Delivery Stream

This sub-command is used to configure the principle parameters of each of the three possible concurrent host delivery streams, one at a time. Following this configuration phase, the particular stream (1, 2, or 3) or all streams (0) may be started, stopped, altered, or undefined, with other sub-commands. This sub-command’s format is:

 

Command

 

 

"c 00 st pppp sync per f num"

 

 

 

 

 

’c’ is the operation code

 

 

 

 

 

’00’ is the sub-command index (ii) for configuration

 

 

 

 

 

’st’ is the stream id digit (1, 2, or 3)

 

 

 

 

 

’pppp’ is a 1-4 hex digit position field (channel

 

 

 

 

 

selection bit map)

 

 

 

 

 

’sync’ is sync type character (0= hardware trigger

 

 

 

 

 

or 1= clock)

 

 

 

 

 

’per’ is the period (if sync=0: # of trigger periods or

 

 

 

 

 

if sync=1: delay timer period in msec).

 

 

 

 

 

’f’ is the format of each acquired datum in stream

 

 

 

 

 

’num’ is the number of scans delivered in the

 

 

 

 

 

stream

 

 

 

 

 

(0=unbounded, >0 bounded).

 

 

 

 

 

NOTE: all parameters are separated by a space

 

 

 

 

 

character.

 

 

 

 

 

 

 

 

Response

 

 

"A" or "Nxx" ’A’ for acknowledge; ’N’ for negative

 

 

 

 

 

acknowledge, error code xx.

 

 

 

 

 

 

 

 

Autonomous

 

 

none generated (until Stream Start sub-command)

 

 

 

 

 

 

Packet

 

 

 

 

 

 

 

 

 

 

Description: Configures a particular stream (‘st’) to deliver data scans autonomously to the host, with each scan containing selected acquired data for the channels specified. The channels are specified by a standard 16-bit position field bit map (encoded pppp as a 1-4 hex digits). A separate sub-command (ii=05) may be used to select which acquired data groups or prefixes are included in each stream. By default, only Temperature/Resistance (EU) data group is selected (if ii=05 sub-command is never executed for the stream after its initializing ‘c 00’ sub- command).

The generation of individual scans of a stream may be synchronized with either an external user-supplied hardware trigger or a periodic (1000 Hz) clock interrupt generated inside each module. This choice is made with the sync type sync’ parameter (a single digit) where: 0 = synchronize with hardware trigger; 1= synchronize with the periodic clock.

Page 44

www.PressureSystems.com

Image 50
Contents Model Ethernet Intelligent Temperature Scanner User’s Manual Page Table of Contents Our Company Our WarrantyMerchandise Return Procedures Technical SupportWebsite and E-Mail Our Firmware Our Publication Disclaimer This page left intentionally blank Introduction IntroductionGeneral Physical Description Exploded View of ModuleTheory of Operation Thermocouple Cold Junction CompensationSetup Instructions Open Circuit DetectionEquipment Required Computer Preparation Network Dialog BoxTCP/IP Properties Dialog Box Module Connection Cable ConnectionsPressure Systems, Inc Model 9046 User’s Manual IP Address Assignment Network Status ScreenNetwork Status after connection ConfigureNetwork Options Pulldown Menu Pressure Systems, Inc Model 9046 User’s Manual Set Sensor Type Sensor Type Values Thermocouple Reference Temperature Alarms Read Sensor TypeC05 Write EU Filter Parameter Set Temperature UnitsRead Temperature Units V0110b ddV00152 Read EU Filter Parameter U00152Calibration U00100 H0001V00100 W08Pressure Systems, Inc Model 9046 User’s Manual Description of Operation Sensor Type Gain Setting Engineering Va, Vb UnitsFirmware Resistance VoltageThermocouple ThermistorUpdating the Firmware This page intentionally blank Command Set Command Quick ReferencePressure Systems, Inc.Model 9046 User’s Manual Summary Information About All Commands Pressure Systems, Inc Model 9046 User’s Manual Pressure Systems, Inc Model 9046 User’s Manual No Operation Command ’A’ Reset Command ’B’ Vppppf Read Primary Transducer Volts Command ’V’Purpose Dddd.. ddddV11110 999999 -4.989500 0.005390Read Primary Transducer A/D Counts Command ’a’ A11110 32767.000000 -32700.000000 10.000000Read Primary Transducer E. U. Data at High Speed Command ’b’ Aaaabbbbcccc..ppppNuss does not use this command normally. You may execute it Define/Control Autonomous Host Streams Command ’c’ Pressure Systems, Inc Model 9046 User’s Manual Pressure Systems, Inc Model 9046 User’s Manual Pressure Systems, Inc Model 9046 User’s Manual Pressure Systems, Inc Model 9046 User’s Manual Read response 00 1 000F 0 1 7 0 c 00 2 00F0 0 2 7 0 c 00 3 FF00 0 4 7 Pressure Systems, Inc Model 9046 User’s Manual Command ’c’- Sub-command Index 01 Start Streams Example Command ’c’- Sub-command Index 02 Stop Streams SpaceExample Command ’c’- Sub-command Index 03 Clear Streams 02 0 c 03 3 c 01Pressure Systems, Inc Model 9046 User’s Manual Command ’c’ Sub-command Index 04 Return Stream Information CharacterDescription Ffff 0 20 7 32000 1 7002 200.200.200.1Prefix/Data Groups Enable UTR Temperature/Other A/D Counts Enable Primary Transducer Temperature/ResistanceEnable UTR Temperature/Other EU Data ûC Enable UTR Temperature/Other Voltages05 1 Command c 06 st pro remport ipaddr 06 0 1 06 0Nuss does not normally use this command since it generates This page intentionally left blank Appendix a Cable Diagrams Model 9046 Ethernet Interface CableModel 9046 Ethernet Interface Cable Headquarters/Factory European Office