Pressure Systems 9016 manual Remove only Perimeter Screws

Page 5

Pressure Systems, Inc.

9016 Upgrade Instructions

Remove 12 perimeter Phillips- head screws.

REMOVE ONLY

THE 12

PERIMETER

SCREWS.

Figure 3

9016 Top Plate

(2)When all screws have been removed, gently lift the top panel and attached electronics up and out of the housing. All components of the pressure scanner are attached to the top plate and will lift out of the module housing when the top plate is removed. See Figure 4, next page. In some cases, it may be easier to hold the top plate and turn the module over, lifting the housing off the top panel. If the Viton® gasket is loose, tack it down with 495 LocTitesuper bonder. (Loctite Corp., USA) (or other suitable adhesive capable of adhering to the gasket material and to the aluminum case).

4www.PressureSystems.com

Image 5
Contents Model Upgrade Instructions Packing List Exploded View of 9016 Scanner Top Plate Upgrade InstructionsUse caution not to slip and scratch the top plate Module DisassemblyRemove only Perimeter Screws PC-206 Board Out of HousingPC-206 Amplifier/Multiplexer Board PC-280 Board PC-280 Board and ConnectorsPC-280 Ethernet Microprocessor/A-D Board Tubing Routing Underneath PC-327 Board Installing the PC-327 Analog BoardPC-322/323 Boards P6 connection PC-327 Board Installing the PC-322/323 BoardsTop Plate to P1 Connection on PC-322 Finding Pin 1 on a PC BoardFigure 9aPC-327 Connection and Cable Tie-downNote Side View of 9016 Upgrade Top View of 9016 UpgradeFinal Checkout Final InstallationHeadquarters/Factory