Pressure Systems 9016 manual PC-327 Connection and Cable Tie-downNote

Page 11

Pressure Systems, Inc.

9016 Upgrade Instructions

(5)Carefully guide the LEDs (on the PC-322 board) into their respective holes in the scanner chassis. Ensure that all cables and tubing are not crimped or restricted and attach the PC- 322/323 boards to the chassis with the three (3) Phillips-head screws through the top plate (that were previously removed and saved). Do not tighten completely.

(6)Connect the ribbon cable from P3 on the PC-322/323 board to the connector on your PC-327 board (red line indicates pin 1). (See Figure 11.)

PC-322/323 connection to the PC-327 board

Figure 11

PC-327 Connection and Cable Tie-downNote

orientation of

red stripe

(7)Insert a plastic cable tie-down through the tie-down mount previously installed and around the clear plastic tubing as shown in Figure 11.

The following two pictures show a completed 9016 module upgrade with side and top views.

Figure 12

Figure 13

Side View of 9016 Upgrade

Top View of 9016 Upgrade

10www.PressureSystems.com

Image 11
Contents Model Upgrade Instructions Packing List Exploded View of 9016 Scanner Top Plate Upgrade InstructionsUse caution not to slip and scratch the top plate Module DisassemblyRemove only Perimeter Screws PC-206 Board Out of HousingPC-206 Amplifier/Multiplexer Board PC-280 Board PC-280 Board and ConnectorsPC-280 Ethernet Microprocessor/A-D Board Tubing Routing Underneath PC-327 Board Installing the PC-327 Analog BoardPC-322/323 Boards P6 connection PC-327 Board Installing the PC-322/323 BoardsTop Plate to P1 Connection on PC-322 Finding Pin 1 on a PC BoardFigure 9aPC-327 Connection and Cable Tie-downNote Side View of 9016 Upgrade Top View of 9016 UpgradeFinal Checkout Final InstallationHeadquarters/Factory