Linak MJB user manual IP Protection degree, Following disinfectants/detergents are recommended

Page 15

IP Protection degree:

All LINAK products can be cleaned as follow according to their IP protection, which is stated on the product label.

The IP code specifies the degrees of protection provided by the enclosures. Only the protection against ingress of water (second characteristic numeral) is specified, ingress of solid foreign objects (first characteristic numeral) is not specified and therefore replaced by the letter X in the code.

IP Protection

Cleaning instrucions

 

 

IPX0

Clean with a damp cloth

 

 

IPX1

Clean with a damp cloth

 

 

IPX2

Clean with a damp cloth

 

 

IPX3

Clean with a damp cloth

 

 

IPX4

Clean with a damp cloth

 

 

IPX5

Wash with a brush and water, but not water under pressure

 

 

IPX6

X Wash with a brush and water. The water can be under pressure, but the system

 

must not be hosed down directly with a high-pressure cleaner. Max. 20oC

IPX6 Washable

Clean by the use of water tunnels

 

 

The materials are resistant to the majority of cleaners and disinfectants used in the hospital and nursing home sector.

Cleaning with a steam cleaner, for example, is not permitted, as it will not be possible to maintain a minimum distance of 300 mm from the electrical parts.

The following directions regarding cleaners and disinfectants must be complied with:

They must not be highly alkaline or acidic (pH value 6-8).

They must not contain caustic agents.

Their contents must not be able to change the structure of the surface or adhesion of the plastic.

They must not break down grease.

The following disinfectants/detergents are recommended:

Neodischer Dekonta, dosage 1-3%, made by CHEMISCHE FABRIK Dr. WEIGERT,

CCOTRADE RW, dosage 0.5%, made by CCOTRADE GmbH.

15

Image 15
Contents MJB Gateway Page Contents Preface Safety instructions Important informationDescription of the various signs used in this manual Residual risk During operation Before installation, re-installation, or troubleshootingInformation on the MJB Before start-upRepairs Disposal of LINAK’s productsComponent Recycling group Warranty and service life MaintenanceMisc. on the MJB system Standard functionality MJB Gateway variant, 504-01Features Options special articleUsage Relay Example of closed loopExample of Open Loop Precautions Cable ordering nos. Modular plug open end IP Protection Cleaning instrucions IP Protection degreeFollowing disinfectants/detergents are recommended Precaution Environmental conditionsEnvironmental conditions Modular plug with narrow slit Modular plug with wide slitKey to symbols ETL LGA MJB Modular Junction Box Drawing AppendixLinak Application Policy Page Page Factories