Simpson Electric H340 manual Temperature Input Cards, Password, Relay Cards

Page 12

6.6 Password

7.TEMPERATURE INPUT CARDS

Pass (Password)

 

 

 

000*

Full Access.

 

001 to 200 No access to scale or factory reset.

 

 

201 to 400

No access to scale, factory reset or

 

measurement control.

 

401 to 600 No access to scale, factory reset,

 

measurement control or output control.

 

 

601 to 800

No access to scale, factory reset,

 

measurement control, output control or relay

RTD

PT100

IN HI

SENSE

IN LO

TEMPERATURE

RTD PT100: The resistance input card has been configured to track the characteristics of a PT100 sensor.

Lead length error can be cancelled out by bringing the sensor to 0 C (ice point) and using the “ICE” command to zero the meter. (Remember to move the ICE capture from the “On” to “Hold” in order to retain the data.)

Lead length error can also be cancelled using the guard wire.

 

 

control.

 

 

 

 

801 to 999 No access except password entry

 

 

 

-999

Keypad lockout. All future key strokes will be

 

 

ignored. All setting changes must come from

 

 

RS-485 port. This choice is available only if

 

 

the RS-485 card is present and configured

 

 

for "Full".

 

 

 

CHG

 

 

 

 

 

Nu-pas

Change the password to a new value. This is

 

(New

available only after successfully entering the

 

Password)

current password or if the password is set to

 

 

000.

 

 

 

NOTE: If you forget your password, turn power off in unit, press and hold the

Tkey and the RESETENTER key. Turn power on and wait until the display reads “noPas” then press the RESET key again to reset password to 000.

ENTER

* Factory Reset default value

7.2J, K, E, T Thermocouple: The thermocouple input card

 

contains all the response information needed to read a J

THERMO

type or K type thermocouple.

COUPLE

Lead length error can be cancelled out by bringing the cen-

J,K,E,T

sor to 0 C (ice point) and using the “ICE” command to zero

 

IN+

the meter. (Remember to move the ICE capture from “On” to

“Hold” to retain the data.

 

IN--

 

!

TEMPERATURE

Never apply a voltage or current to these inputs as damage will occur.

7.1Relay Cards

Single Relay

Dual Relay

Quad Relay

N.C.

1 COM

N.O.

250V~ 5A

RELAY OUT

12

 

N.C.

 

1

COM

 

 

N.O.

 

250V~

5A

 

N.C.

 

2

COM

 

 

N.O.

 

RELAY OUT

 

 

13

 

N.O.

1

COM

N.O.

2

COM

N.O.

3

COM

4

N.O.

RELAY OUT

250V 5A

Image 12
Contents Operators Manual Simpson Electric Company 520 Simpson AvenueWarranty and Returns Technical AssistancePage Contents Specifications IntroductionOutput Control General DescriptionInstallation and Panel Cutout Display and Keypad ControlsDisplay Control Mounting RequirementsEngineering Label Placement Menu DetailRelay Control Sp1-Sp4 Run Mode Program ModeEdit Mode Operating the KeysMenu Flowchart Programming Menu Relay ControlMeasurement Control Frequently Asked Questions Power Supply CardOutput Control If Present System ControlTemperature Input Cards PasswordRelay Cards Single Relay Dual Relay Quad Relay