6.6 Password |
7.TEMPERATURE INPUT CARDS
Pass (Password) |
|
|
|
000* | Full Access. |
| |
001 to 200 No access to scale or factory reset. | |
|
|
201 to 400 | No access to scale, factory reset or |
| measurement control. |
| |
401 to 600 No access to scale, factory reset, | |
| measurement control or output control. |
|
|
601 to 800 | No access to scale, factory reset, |
| measurement control, output control or relay |
RTD
PT100
IN HI
SENSE
IN LO
TEMPERATURE
RTD PT100: The resistance input card has been configured to track the characteristics of a PT100 sensor.
Lead length error can be cancelled out by bringing the sensor to 0 C (ice point) and using the “ICE” command to zero the meter. (Remember to move the ICE capture from the “On” to “Hold” in order to retain the data.)
Lead length error can also be cancelled using the guard wire.
|
| control. |
|
|
|
| 801 to 999 No access except password entry | |
|
|
|
Keypad lockout. All future key strokes will be | ||
|
| ignored. All setting changes must come from |
|
| |
|
| the |
|
| for "Full". |
|
|
|
CHG |
| |
|
|
|
| Change the password to a new value. This is | |
| (New | available only after successfully entering the |
| Password) | current password or if the password is set to |
|
| 000. |
|
|
|
NOTE: If you forget your password, turn power off in unit, press and hold the
Tkey and the RESETENTER key. Turn power on and wait until the display reads “noPas” then press the RESET key again to reset password to 000.
ENTER
* Factory Reset default value
7.2J, K, E, T Thermocouple: The thermocouple input card
| contains all the response information needed to read a J | |
THERMO | type or K type thermocouple. | |
COUPLE | Lead length error can be cancelled out by bringing the cen- | |
J,K,E,T | sor to 0 C (ice point) and using the “ICE” command to zero | |
| ||
IN+ | the meter. (Remember to move the ICE capture from “On” to | |
“Hold” to retain the data. | ||
| ||
|
!
TEMPERATURE
Never apply a voltage or current to these inputs as damage will occur.
7.1Relay Cards
Single Relay | Dual Relay | Quad Relay |
N.C.
1 COM
N.O.
250V~ 5A
RELAY OUT
12
| N.C. |
|
1 | COM |
|
| N.O. |
|
250V~ | 5A | |
| N.C. |
|
2 | COM |
|
| N.O. |
|
RELAY OUT |
| |
| 13 |
|
N.O.
1
COM
N.O.
2
COM
N.O.
3
COM
4
N.O.
RELAY OUT
250V 5A